Model 407764 Version 1.4  December 2004 

2

 

Specifications 

Applicable Standards

  

IEC651 Type 2, ANSI S1.4 Type 2 

Accuracy

 

±1.5dB (under reference conditions) 

Frequency range

  

31.5Hz - 8KHz 

Measuring level

  

30 - 130dB 

Frequency weighting

  

A and C 

Microphone

 

0.5” Electret condenser microphone 

Display

 4-digit 

LCD 

 Resolution: 

0.1dB 

 

Display period: 0.5 sec. 

Bargraph

 

50dB scale (1dB steps). Display period: 50mS; Auto-ranging: 
100dB scale, 2dB steps 

Sampling rate

 50mS

  

Memory size

 

16,000 records / 255 sets (non-volatile memory) 

Datalogging sample rate 

1 to 86,400 seconds per record

 

Time weighting

 

FAST: 125mS, SLOW: 1 sec. 

MAX

 

Maximum reading held 

Level ranges

  

30-80dB, 40-90dB, 50-100dB, 60-110 dB, 70-120dB, 80-130dB 
(Total of 6 ranges) 

Auto range

  

30 to 130dB 

Linearity range

  

50dB 

Alarm function

  

 

OVER indicator for readings higher than high limit. 

 

UNDER indicator for readings lower than low limit. 

AC output

    

0.707 Vrms at Full Scale 

 

Output impedance approx. 600

 

DC output

 

10mV / dB 

 

Output impedance approx. 100

 

Power supply

  

Four 1.5V ‘AA’ batteries 

Battery life 

 

Approx. 30 hrs continuous operation 

AC adapter

  

Voltage: 6VDC 

 

Voltage Ripple: < 100mVpp 

 

Supply Current: > 100mADC 

 

Socket: Pin Ground 

 Casing: 

Positive 

 

External Diameter: 3.5mm 

Operating temperature

  

32 to 104

o

F (0 to 40

o

C) 

Operating humidity

  

10 to 80%RH 

Storage temperature

  

14 to 140

o

F (-10 to 60

o

C) 

Storage humidity

  

10 to 80%RH 

Memory size

  

16,000 data records with Date and Time Stamping 

RS-232 Interface  

Baud rate: 9600bps 

Dimensions

  

10.4 x 2.8 x 1.4” (265 × 72 × 35mm 

Weight

  

Approx. 11.5 oz. (358g) including battery 

Accessories

  

Batteries, carrying case, screwdriver, windscreen, 3.5mm plug, RS-232 
cable , 9- to 25-pin gender changer, Windows

TM

  compatible software. 

 

Safety symbols 

Meter is protected throughout by double insulation or reinforced 
insulation. 

CE

  

Complies with EMC 

Содержание 407764

Страница 1: ...atulations on your purchase of Extech s Datalogging Sound Level Meter The meter is capable of performing all your noise measurements needs in addition to it s capabilities for data acquisition direct data storage to a PC or datalogging data storage to internal memory for later download Data storage enables the user to save data analyze data and generate reports This professional meter with proper ...

Страница 2: ...s higher than high limit UNDER indicator for readings lower than low limit AC output 0 707 Vrms at Full Scale Output impedance approx 600Ω DC output 10mV dB Output impedance approx 100Ω Power supply Four 1 5V AA batteries Battery life Approx 30 hrs continuous operation AC adapter Voltage 6VDC Voltage Ripple 100mVpp Supply Current 100mADC Socket Pin Ground Casing Positive External Diameter 3 5mm Op...

Страница 3: ...1 External DC 6V power 12 RS 232 interface connector 13 Battery cover 14 Tripod mounting screw 15 Wind screen Figure 1 Display Description 1 Range selection 2 Date information 3 SPL Instantaneous sound pressure level 4 Low Battery 5 MAX Maximum SPL value is held 6 Data recording 7 Measurement value 8 Memory full 9 Units 10 Frequency weighting A C 11 TIME function 12 50dB level Bargraph 13 Over ran...

Страница 4: ...e Frequency Weighting Change the Frequency Weighting by pressing the A C button The A or C icon will display on the right hand area of the LCD Note With A weighting selected the meter responds like the human ear boosting and cutting the noise amplitude over the frequency spectrum see Appendix A weighting is used for environmental measurements OSHA regulatory testing law enforcement and workplace d...

Страница 5: ...adjust the Sound Level Meter s CAL potentiometer as shown in figure at right to match the Calibrator s output 4 Typically a Sound Level Calibrator will supply 94dB or 114dB at 1KHz Adjust the Sound Level Meter s CAL potentiometer for a 94dB or 114dB LCD display Figure 3 Analog output The 407764 has two analog output jacks located on its right side one for AC and one for DC For DC the meter transmi...

Страница 6: ... cleared 6 To Download or Record data via a PC refer to the PC Interface section below PC Interface Connecting the Meter to a PC Refer to Figure 5 Connect the 9 pin male connector to the Sound Level Meter and connect the 9 pin female connector to the 9 pin COM1 PC port If COM1 is used by the mouse connect the 25 pin female connector to the 25 pin COM2 PC port you ll need to use the included 9 to 2...

Страница 7: ... Displays the Real Time Clock Setting SAMPLING TIME Display of data acquisition sampling rate Fig 7 PC SAMPLING Opens data acquisition sampling rate selection box Datalogging sample rate is set in the MEMORY Logger window RESET Resets the MIN and MAX stored readings SYSTEM TIME SET Sets the meter s clock to the value of the pc clock SAVE AS Opens the data file storage window dat OPEN FILE Opens th...

Страница 8: ...licked downloads and displays the meter s date and time ID CODE Numeric Identification code Enter the code in the box and click on the bar to store the code SAMPLING Datalogger sample rate Enter the sample rate in the box in seconds and click on the SAMPLING bar to store the value NUMBER OF SETS Click on the bar to download the stored data TIME OF RECORDING Click on the bar to display sets downloa...

Страница 9: ...ata will be downloaded from meter to PC Save to disk if desired 5 Click the SHOW DATA button for details on each record Figure 13 shows the data list Select SAVE PRINT GRAPH or EXIT from the menu choices Note In the example for Figure 12 there are 3 sets of recorded data in memory Set No 3 has 15 records Note If you need to change the ID CODE or the SAMPLING time click on the desired parameter typ...

Страница 10: ...tment at 781 890 7440 ext 210 for authorization or visit our website at www extech com click on Contact Extech and go to Service Department to request an RA number A Return Authorization RA number must be issued before any product is returned to Extech The sender is responsible for shipping charges freight insurance and proper packaging to prevent damage in transit This warranty does not apply to ...

Страница 11: ...80 60 40 20 0 130 110 90 70 50 30 10 50 HP Siren 100 Jet take off 200 Riveting machine Subway 20 Pneumatic drill 50 Vacuum cleaner 10 Large store Small office Residential area night Whisper 5 Threshold of hearing Chain saw Boiler room Freight train 100 Speech 1 Large Office Residence Sound studio ...

Страница 12: ...yte year month date hour min sec Set sampling rate Send 3 bytes W high byte low byte leading byte 1 65535 sec Set ID code Send 3 bytes X high byte low byte leading byte 1 65535 Get software version Output V Receive 1 byte 0 255 Get system argument Output Y Receive 10 bytes ID Code 2bytes Version 1 byte ID code 1 byte Data sets 1 byte Last address 3 bytes Sample rate 2 bytes Get time and date Outpu...

Отзывы: