Extech Instruments 407750 Скачать руководство пользователя страница 8

Model 407750 Version 2.1 February 2004 

8

 

Support Hotline (781) 890-7440 

Tech support: Ext. 200; Email: [email protected] 

Repair/Returns: Ext. 210; Email: [email protected] 

Website: www.extech.com 

Battery Replacement 

When the low battery message appears on the LCD, the 9V battery has fallen to a critically 
low voltage level and should be replaced as soon as possible. The battery compartment 
cover resides at the rear of the meter. Remove the rear battery compartment screw and 
remove the battery compartment cover, change the battery, and replace the compartment 
cover. 

Calibration and Repair 

The meter includes a Calibration adjustment on the bottom of the meter. A sound level 
calibrator must be used to complete a calibration. Extech can supply a calibrator for this 
purpose. Once the Calibrator is affixed on the microphone, adjust the calibration screw so 
that the LCD sound level display matches the calibrator’s output signal.  
 
Extech offers repair and calibration services for the products we sell. Extech also provides 
NIST certification for most products. Call the Customer Service Department for information 
on calibration services available for this product. Extech recommends that annual 
calibrations be performed to verify meter performance and accuracy. 
 

 
 
 
 
 
 
 
 
 

 

Copyright 

©

 2004 Extech Instruments Corporation 

All rights reserved including the right of reproduction in whole or in part in any form. 

 
 
 

Содержание 407750

Страница 1: ...User s Guide Digital Sound Level Meter Model 407750 ...

Страница 2: ...d to repair or replacement of the product The warranty set forth above is inclusive and no other warranty whether written or oral is expressed or implied Introduction Congratulations on your purchase of the Extech 407750 This device measures sound level in dB and the measurement range can be set automatically or manually The 407750 offers selectable frequency weighting A and C and Time Response Fa...

Страница 3: ...ct key 7 BA Background Absorber key 8 F S Fast Slow Response select key 9 DOWN 10 Backlit LCD backlighting key 11 UPPER 12 AC adaptor jack 13 Calibration screw adjust 14 AC analog output jack 15 DC analog output jack 16 RS 232 output jack Note The Battery Compartment and the Threaded Tripod mount access are located on the back of the instrument not pictured ...

Страница 4: ...OW 1 second response measurements by pressing the F S key The LCD will reflect the currently selected mode Selection of Fast or Slow is determined by the application and any directives or standards related to that application For example most hearing conservation or OSHA related testing is done using SLOW and A weighting MAX HOLD The meter is capable of taking continuous measurements and only upda...

Страница 5: ...ally finds the correct range in order to produce the best accuracy However if it is desired to set the range manually follow these steps 1 Power the meter 2 Notice the two 2 digit number to the immediate left of the analog bargraph This number is the low end of the presently selected range see the specifications for the ranges 3 To change the range press the UP key to raise the range or press the ...

Страница 6: ...AC output is 0 707V rms full scale and the DC output is 10mV per dB The labeled 3 5mm output mini plugs are located on the bottom of the instrument RS 232 Output The meter includes an RS 232 PC interface jack This PC interface allows the meter to store and display readings on a PC as they are recorded The interface cable and software for data acquisition are sold separately Detailed instructions a...

Страница 7: ...5dB 500 3 2dB 0dB 1 5dB 1 k 0dB 0dB 1 5dB 2 k 1 2dB 0 2dB 2dB 4 k 1dB 0 8dB 3dB 8 k 1 1dB 3dB 5dB Typical A Weighted Sound Levels 50HP Siren 100 130 120 110 100 90 80 70 60 50 40 30 20 10 140 0 Jet takeoff 200 dB Riveting machine Chain saw Subway 20 Boiler room Pneumatic drill 50 Freight train 100 Vacuum cleaner 10 Speech 1 Large store Large Office Small office Residence Night residential area Whi...

Страница 8: ... includes a Calibration adjustment on the bottom of the meter A sound level calibrator must be used to complete a calibration Extech can supply a calibrator for this purpose Once the Calibrator is affixed on the microphone adjust the calibration screw so that the LCD sound level display matches the calibrator s output signal Extech offers repair and calibration services for the products we sell Ex...