background image

Model 407740 version 1.6  8/05 

Battery Replacement 

The low battery indication (BAT) appears on the LCD when the 9V battery is nearing 
depletion. Reliable readings can be obtained for several hours after the first appearance of 
the low battery indication. To replace the battery, slide the battery compartment cover off 
the rear of the instrument and replace the 9V battery. Secure the cover before use. 

Reference Information 

Frequency Weighting Characteristics 

 

Frequency (Hz)

 

A Weighting

 

C Weighting

 

Tolerance

 

(IEC 651 Type 2

31.5 -39.4dB 

-3dB 

±3dB 

63 -26.2dB 

-0.8dB 

±2dB 

125 -16.1dB 

-0.2dB 

±1.5dB 

250 -8.6dB 

0dB 

±1.5dB 

500 -3.2dB 

0dB 

±1.5dB 

1 k 

0dB 

0dB 

±1.5dB 

2 k 

+1.2dB 

-0.2dB 

±2dB 

4 k 

+1dB 

-0.8dB 

±3dB 

8 k 

-1.1dB 

-3dB 

±5dB 

Typical A-Weighted Sound Levels 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 

 

 

 
 
 

 
 

Calibration and Repair Services 

Extech offers repair and calibration services for the products we sell. Extech also provides 
NIST certification for most products. Call the Customer Service Department for information 
on calibration services available for this product. Extech recommends that annual 
calibrations be performed to verify meter performance and accuracy. 

 
 

 

Copyright 

©

 2003 Extech Instruments Corporation 

All rights reserved including the right of reproduction in whole or in part in any form. 

50HP Siren (100')

130

120

110

100

90

80

70

60

50

40

30

20

10

140

0

Jet takeoff (200')

dB

Riveting machine

 Chain saw

Subway (20')

Boiler room

Pneumatic drill (50')

 Freight train (100')

Vacuum cleaner (10')

Speech (1')

Large store

Large Office

Small office

 Residence

Night residential area

Whisper (5'0)

 Sound studio

Threshold of hearing

 

Содержание 407740

Страница 1: ...gratulations on your purchase of the Extech 407740 Digital Sound Level Meter This device measures sound in decibels dB Useful features include A C Weighting Fast Slow Response Time and Maximum Reading HOLD Careful use of this meter will provide years of reliable service ...

Страница 2: ... improper maintenance or repair or unauthorized modification Extech specifically disclaims any implied warranties or merchantability or fitness for a specific purpose and will not be liable for any direct indirect incidental or consequential damages Extech s total liability is limited to repair or replacement of the product The warranty set forth above is inclusive and no other warranty whether wr...

Страница 3: ...orcement and workplace design Select C weighting for flat response measurements the readout does not boost or dampen the measurement across the frequency spectrum C weighting is suitable for the operational maintenance and analysis of machinery motors pumps engines etc Selecting FAST or SLOW response Use the RESPONSE switch to select Fast or Slow The setting is determined by the application and an...

Страница 4: ...0 8dB 3dB 8 k 1 1dB 3dB 5dB Typical A Weighted Sound Levels Calibration and Repair Services Extech offers repair and calibration services for the products we sell Extech also provides NIST certification for most products Call the Customer Service Department for information on calibration services available for this product Extech recommends that annual calibrations be performed to verify meter per...

Отзывы: