Extech Instruments 407730 Скачать руководство пользователя страница 5

Model 407730 Version 1.3 June 2004 

5

Specifications 

Display LCD 

with 

bargraph

 

Microphone 

0.5” (10mm) Electret condensor 

Measurement Bandwidth 

300Hz to 8KHz

 

Measurement Range 

40 to 130dB (A wtg), 45 to 130dB (C wtg) 

Frequency weighting 

‘A’ and ‘C’ (selectable) 

Accuracy / Resolution 

± 2dB (under reference conditions) / 0.1dB 

Response time 

Fast: 125 milliseconds / Slow: 1 second 

Calibration source 

1KHz sine wave @ 94 or 114dB 

AC output 

0.707Vrms full scale 

Power 

4 AAA Batteries

 

Battery life 

30 hours (typical); low battery indicator alerts user 

Automatic power off 

After approx. 20 minutes  

Operating temperature 

32 to 122

o

F (0 to 50

o

C) 

Operating humidity 

10 to 90% RH 

Storage temperature 

-4 to 140

o

F (-20 to 60

o

C) 

Dimensions/weight 

9 x 2.3 x 1.7” (230 x 57 x 44mm) / 6oz (172g)

 

Warranty 

EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects 
in parts and workmanship for one year from date of shipment (a six month limited warranty 
applies to sensors and cables). If it should become necessary to return the instrument for 
service during or beyond the warranty period, contact the Customer Service Department at 
(781) 890-7440 ext. 210 for authorization or visit our website www.extech.com for contact 
information. A Return Authorization (RA) number must be issued before any product is 
returned to Extech. The sender is responsible for shipping charges, freight, insurance and 
proper packaging to prevent damage in transit. This warranty does not apply to defects 
resulting from action of the user such as misuse, improper wiring, operation outside of 
specification, improper maintenance or repair, or unauthorized modification. Extech 
specifically disclaims any implied warranties or merchantability or fitness for a specific 
purpose and will not be liable for any direct, indirect, incidental or consequential damages. 
Extech's total liability is limited to repair or replacement of the product. The warranty set 
forth above is inclusive and no other warranty, whether written or oral, is expressed or 
implied. 

Calibration and Repair Services 

Extech offers repair and calibration services

 for the products we sell. Extech also 

provides NIST certification for most products. Call the Customer Service Department for 
information on calibration services available for this product. Extech recommends that 
annual calibrations be performed to verify meter performance and accuracy. 

Содержание 407730

Страница 1: ...User s Guide Digital Sound Level Meter Model 407730 ...

Страница 2: ...esponse Time Fast and Slow Max Hold and Max Min recording Careful use of this meter will provide years of reliable service Meter Description 1 Microphone 2 LCD Display 3 ON OFF button 4 A C weighting selection button 5 Min Max Record button 6 Range selector button 7 F S response selection button 8 Max Hold selector button 9 Calibration adjustment 10 AC analog output jack 11 Windscreen 1 Record ico...

Страница 3: ...Time Use the F S button to select FAST 125 ms or SLOW 1 second response time Select FAST to capture noise peaks and noises that occur very quickly Select the SLOW response to monitor a sound source that has a consistent noise level or to average quickly changing levels FAST or SLOW icons will appear in the display Select Slow response for most applications Auto or Manual Ranging The meter will tur...

Страница 4: ...button is pressed Calibration To calibrate the meter an external calibrator such as the Extech 407744 or the Extech 407766 is required in addition to a small screw driver 1 Turn the meter ON 2 Select the 80 to 110dB range 3 Select A weighting and SLOW response 4 Place the microphone into the calibrator Set the calibrator to output a 1kHz sine wave 94dB 5 Adjust the calibration potentiometer for a ...

Страница 5: ...tomer Service Department at 781 890 7440 ext 210 for authorization or visit our website www extech com for contact information A Return Authorization RA number must be issued before any product is returned to Extech The sender is responsible for shipping charges freight insurance and proper packaging to prevent damage in transit This warranty does not apply to defects resulting from action of the ...

Страница 6: ...t information visit our website www extech com Typical A Weighted Sound Levels 50HP Siren 100 130 120 110 100 90 80 70 60 50 40 30 20 10 140 0 Jet takeoff 200 dBA Riveting machine Chain saw Subway 20 Boiler room Pneumatic drill 50 Freight train 100 Vacuum cleaner 10 Speech 1 Large store Large Office Small office Residence Night residential area Whisper 5 0 Sound studio Threshold of hearing Copyrig...

Отзывы: