background image

User's Guide 

 
 
 
 
Heavy Duty Conductivity / Temperature 
 
Model 407303

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Introduction 

Congratulations on your purchase of Extech’s Conductivity/ Temperature Meter. This 
meter offers a dual display for Conductivity and Temperature, a Conductivity Cell 
calibration adjustment, temperature compensation, and a RS-232 PC Interface. This meter 
is shipped fully tested and calibrated and with proper use will provide years of reliable 
service. 

 
 

 

 

99 Washington Street 
Melrose, MA 02176 

Phone

 781-665-

1400

Toll Free 1-800-517-8431

 

Visit us at www.T

estEquipmentDepot.com

Back to the 

Extech 407303

 Product Page

Содержание 407303

Страница 1: ...onductivity and Temperature a Conductivity Cell calibration adjustment temperature compensation and a RS 232 PC Interface This meter is shipped fully tested and calibrated and with proper use will provide years of reliable service 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com Back to the Extech 407303 Product Page ...

Страница 2: ...ings for later recall Auto Power OFF Automatic power off after 10 minutes PC Interface RS 232 serial data output Operating conditions Temperature 32 F to 122 F 0 C to 50 C 80 R H Power Supply 9V battery Power Consumption Approx 10 mA DC Weight 17 oz 482 g including batteries probe Dimensions Main instrument 7 1 x 2 8 x 1 3 180 x 72 x 32 mm Probe 0 87 22 mm diameter Measurement Range Resolution Acc...

Страница 3: ... the displayed reading C F Press to select the temperature unit of measure RECORD Press to begin tracking the MIN MAX readings RECALL Press to retrieve the MIN MAX readings FACTOR ADJUST Used with the TEMP button for temperature compensation RANGE Slide switch to manually select the measurement range TEMP Temperature coefficient compensation button 5 Battery compartment rear Remove protective hols...

Страница 4: ... temperature readings will appear on the LCD Conductivity measurements appear at the center of the LCD Temperature readings appear on the bottom of the LCD in smaller digits 3 Allow time for the measurements to stabilize The LCD display digits will settle after a short stabilization period 4 If dashes appear on the top of the LCD measurement area the reading is over range Slide the RANGE select sw...

Страница 5: ...n The REC MAX MIN and AVG icons should be off when the meter has exited the MIN MAX AVG Record Recall mode Conductivity Calibration 1 Keep the battery connected to the meter but remove it from the battery compartment 2 Prepare a standard conductivity solution of one or all of the following 200uS 2 000mS and 20 00mS 3 Place the probe into the calibration solution and adjust the corresponding potent...

Страница 6: ...ed at the top of the meter next to the sensor input jack The interface is intended for use with the data acquisition software and hardware kit For more information contact Extech or refer to the Software User Manual Battery Replacement The low battery indicator LBT appears on the LCD when the battery runs low To replace the battery 1 Remove the meter s rubber protective cover to access the rear ba...

Отзывы: