407026-EU-EN v2.6 3/14
2
Introduction
Congratulations on your purchase of the Extech Heavy Duty Light Meter. This light meter offers
selectable lighting types, data record/recall, relative display mode, and PC interface. This
professional meter, with proper care, will provide years of safe reliable service.
Specifications
General Specifications
Circuit
Custom one-chip LSI microprocessor circuit
Display
3-1/2 digit (2000 count) LCD display with contrast adjustment
Measurement ranges
LUX: 0 to 50,000 LUX (3 range); Fc: 0 to 5000 Fc (3 range);
Relative mode: 0 to 1999%
Data Hold
Freezes displayed reading
Lighting Types
Sodium, Daylight/Tungsten, Fluorescent and Mercury
Sensor Structure
Cosine/color corrected photo-diode meets C.I.E.
Memory Store/Recall
Records/Recalls Max/Min/Avg readings
Sample rate
0.4 seconds (approx.) per reading
Zero Adjust
Push-button procedure
Auto Power Off
After approx. 10 minutes
Data Output
RS-232 PC serial interface (optional software/cable PN 407001)
Operating conditions
0 °C to 50 °C (32 °F to 122 °F); <80% RH
Power Supply
9V battery
Power consumption
Approx. 5 mA DC. (approx. 200 hr battery life)
Weight
320 g (0.71 lbs.)
Dimensions
Instrument: 180 x 72 x 32 mm (7.1 x 2.8 x1.3")
Sensor: 85 x 55 x 17.5 mm (3.3 x 2.2 x 0.7")
Range Specifications
Measurement
Range
Display
Resolution
Accuracy
LUX
2,000 LUX
0-1,999 LUX
1 LUX
± (4% + 2 digits)
of full scale
20,000 LUX
1,800-19,990 LUX
10 LUX
50,000 LUX
18,000-50,000 LUX
100 LUX
Foot Candles
200 Fc
0-186.0 Fc
0.1 Fc
2,000
Fc
167-1,860 Fc
1 Fc
5,000
Fc
1,670-5,000 Fc
10 Fc
Relative mode
0-1999% 1%
Note:
The accuracy specification above applies to calibration performed using a precision standard
incandescent tungsten light source of 2856
O
K with meter on the tungsten setting.
Содержание 407026