382260-EN V1.5   7/12 

Trouble Shooting 

PROBLEM INDICATIONS POSSIBLE 

CAUSES 

SUGGESTED 

SOLUTIONS 

Power supply not 
working 

Panel display, LED 
Indicators not on 

A.  AC power input not 

connected 

B. AC 

input 

fuse 

blown 

A. Check 

AC 

power 

connection, 

B. Contact 

factory 

No DC output power 

Voltage meter zero 
indication. 
ALM (13) Alarm led on 
CC (12) led on 

A.  Output on off button 

not on . 

B.  UVL (10 ) 

protection triggered 

C. OTP 

protection 

triggered 

D. Output 

short 

circuit 

A.  Check output LED 

(9) is on or not, 
push (9) to on. 

B.  C h e c k   U V L   s e t  

voltage by pushing 
(6), re-set UVL to 
applicable limit. 

C. Check vent holes at 

top and bottom is 
clear, ambient 
temperature too 
high  

D. Check and undo 

short circuit of 
output connection. 

Voltage meter reading 
grossly inaccurate 

Actual output voltage is 
grossly different from 
the meter read out 

A. Volt 

meter 

shows 

not the output 
voltage. 

B. Possible 

misalignment in 
voltmeter 
calibration 

A .   C h e c k   t h e   L E D  

indicator of UVL on 
the top right hand 
side of the Volt 
meter, if it lights up 
then the volt 
meter only shows 
the set UVL value. 
Push (6) to go 
back to output 
voltage. 

B .  Contact  factory 

Ampere meter reading 
grossly inaccurate 

In CV mode , the actual 
current measured is 
grossly different from 
the Amp meter reading 

Possible misalignment 
in calibration . 

Contact factory 

The activated 
(auto- cross  over ) 
current limiting value 
is different from the 
preset cc value 

All indicators and 
display are  normal  , 
only CC mode has the 
problem 

For stand alone unit, 
Master & Slave 
Switch (17) in wrong 
position (Slave). Wrong 
setting or connection in 
Master  &  S l a v e   o r  
s e t  u p  procedure. 

Check the Master & 
Slave switch (17) 
correct position. Use 
stand alone mode to 
check power supply 
separately without any 
connection to slave 
u n i t s .   F o l l o w   t h e  
procedure carefully and 
make sure there is only 
one Master unit. 

 

 