EXFO T200S Скачать руководство пользователя страница 83

SCPI Command Reference

Tunable Laser Source

77

General Commands

*SRE

Description

The *SRE command sets the Service Request Enable Register bit 

value, as defined in 

Service Request Enable Register (SRE)

 on 

page 47.

Type

Overlapping.

Syntax

*SRE<wsp><register value>

Parameter(s)

register value: 

The <register value>, represents the bit values of the Service Request 

Enable Register. 
The <register value> value ranges from 0 through 255.
See the content of this register in 

Service Request Enable Register 

(SRE)

 on page 47.

Example(s)

*SRE 52

where 52 sets the ESB, MAV and EAV bits

See Also

*SRE?

 on page 77

*STB?

 on page 78

*SRE?

Description

The *SRE? query returns the current content of the Service 

Request Enable Register. 
See the content of this register in 

Service Request Enable Register 

(SRE)

 on page 47.

Type

Overlapping.

Syntax

*SRE?

Parameter(s)

None.

Response Syntax

<register value>

Response(s)

register value: 

The <register value> represents the current bit values of the 

Service Request Enable Register.
The <register value> ranges from 0 through 255.
See the content of this register in 

Service Request Enable Register 

(SRE)

 on page 47.

Example(s)

*SRE returns 32 (the ESB bit is set)

See Also

*SRE?

 on page 77

*STB?

 on page 78

Содержание T200S

Страница 1: ...User Guide T200S Tunable Laser Source www EXFO com...

Страница 2: ...that may result from its use No license is granted by implication or otherwise under any patent rights of EXFO EXFO s Commerce And Government Entities CAGE code under the North Atlantic Treaty Organiz...

Страница 3: ...e Date and Time 20 Connecting the T200S to your Company Network 21 Renaming the Instrument 22 Handling User Configuration 23 5 Setting the Emission Wavelength Frequency 25 Defining General Emission Se...

Страница 4: ...ing Assistance Tools 64 Viewing System Information 65 Displaying the User Documentation 65 Contacting the Technical Support Group 65 Transportation 66 10 Warranty 67 General Information 67 Gray Market...

Страница 5: ...l et utilis conform ment la documentation de l utilisateur il peut occasionner une interf rence n faste aux communications radio L utilisation de cet quipement dans une zone r sidentielle est suscepti...

Страница 6: ...et de laboratoire les processus industriels appareils de mesure ou de contr le Appareils du groupe 2 le groupe 2 r unit tous les appareils ISM fr quences radio lectriques dans lesquels de l nergie fr...

Страница 7: ...most recent technical specifications visit the EXFO Web site at www exfo com Hardware Interfaces for External Devices USB A 2 0 x2 Mouse keyboard or USB key for software update USB A 3 0 x1 Remote Int...

Страница 8: ...mote examples if any and user documentation Front panel On Off button The On Off button enables you to turn on or off the T200S see Turning on the T200S and Accessing the GUI on page 16 andTurning off...

Страница 9: ...details see Triggering the Sweep Sequence on page 34 Trig out output BNC connector for outputting a trigger signal during wavelength frequency emission For more details see Generating an Output Trigg...

Страница 10: ...For more details see Connecting the T200S to your Company Network on page 21 Perform remote control operations on the T200S For more details see Remotely Controlling the T200S on page 43 Cooling Fan T...

Страница 11: ...egulatory Information on page v UK conformity assessment See Regulatory Information on page v CE marking See Regulatory Information on page v Laser safety label The product is a class 1M laser product...

Страница 12: ...roceed unless you understand and meet the required conditions CAUTION Indicates a potentially hazardous situation which if not avoided may result in minor or moderate injury Do not proceed unless you...

Страница 13: ...d approved by EXFO For a complete list of accessories available for your unit refer to its technical specifications or contact EXFO IMPORTANT Refer to the documentation provided by the manufacturers o...

Страница 14: ...your product are located throughout this documentation depending on the action to perform Make sure to read them carefully when they apply to your situation WARNING Viewing the laser output with tele...

Страница 15: ...or more of the following symbols may also appear on your unit Symbol Meaning Direct current Alternating current The unit is equipped with an earth ground terminal The unit is equipped with a protectiv...

Страница 16: ...NING Use this unit indoors only Do not remove unit covers during operation Operation of any electrical instrument around flammable gases or fumes constitutes a major safety hazard To avoid electrical...

Страница 17: ...formance guaranteed temperature 18 C to 28 C 64 4 F to 82 4 F Relative humidity 80 non condensing Maximum operation altitude 2000 m 6562 ft Pollution degree 2 Overvoltage category II Measurement categ...

Страница 18: ......

Страница 19: ...the T200S on a flat stable surface free of excessive vibration 4 Allow the flow of air to circulate freely at the front and rear of the T200S and remove any equipment or paper that could block the air...

Страница 20: ...e rear panel of the T200S 3 Plug the other end of the power supply cable to the proper voltage wall socket outlet to know the voltage requirement see Electrical Safety Information on page 10 4 On the...

Страница 21: ...nt panel lift the protective cap and connect an external optical component or instruments to the T200S via the FC APC fiber optic connector To protect the laser output connector leave a fiber optic ca...

Страница 22: ...n during startup in this case you can fully use the T200S but performances are not guaranteed The touch screen display enables you to access all the T200S functions To access the T200S graphical user...

Страница 23: ...page 25 Opens the T200S settings window to define general parameters connect the T200S update the T200S or perform wavelength reference Opens the event log For more details see Displaying the List of...

Страница 24: ...perating mode Appears if the T200S needs internal referencing This button performs an internal wavelength referencing For more details see Performing Internal Wavelength Referencing on page 37 Appears...

Страница 25: ...y Colors You can change the theme of the display and modify its brightness as explained in the following procedures To modify the brightness and theme 1 In the T200S task bar touch the button to displ...

Страница 26: ...nd time area enables you to set the date and time of the T200S 2 To manually set the date and time 2a Clear the Set time automatically check box 2b Touch the Date and Time fields to enter the date and...

Страница 27: ...y network 1 Connect an RJ45 cable from your company network to the T200S Ethernet port 2 In the T200S task bar touch the button to display the Settings window and select the Ethernet tab The TCP socke...

Страница 28: ...automatically established You cannot modify the connection values 6 Touch Apply to validate the set connection parameters 7 If you want to retrieve the previously applied connection parameters touch t...

Страница 29: ...button to display the Settings window and select the Instrument tab The Configuration area enables you to manage your laser configuration 2 Select the Keep configuration after reboot check box The la...

Страница 30: ...the task bar touch the button to display the Settings window and select the Instrument tab The Configuration area enables you to manage your laser configuration 2 Touch the Restore factory settings bu...

Страница 31: ...Spectral Unit You can define the spectral unit you want to use for all operating modes in the Settings window as explained in the following procedure To set the spectral unit 1 In the T200S task bar...

Страница 32: ...s to the absolute wavelength or frequency position with optimum absolute accuracy according to your laser optical specifications Activating this setting may slightly slow down the move of the laser to...

Страница 33: ...equency is reached with optimum accuracy To enter the emission wavelength frequency 1 In the task bar select the TUNE mode in the drop down list 2 Enable the laser output see Enabling Disabling the La...

Страница 34: ...es Tuning the laser using steps reached relatively from the absolute wavelength frequency position cancels high accuracy In Optimized linewidth mode the steps are set to 0 005 nm and 0 001 nm or 0 5 G...

Страница 35: ...ngth frequency position Generating an Output Trigger Signal during Emission In TUNE mode the T200S can generate a trigger signal once the laser has stabilized on the selected position for more details...

Страница 36: ...stop wavelength according to the parameters speed pause cycles you have set You can start the sweep manually or with a trigger signal Configuring the Continuous Sweep The Sweep mode settings menu ena...

Страница 37: ...omatically positioned at the start value and moves to the stop value Cycles Number of sweeps you want the laser to perform If you enter 0 the laser sweeps indefinitely until you touch the Stop button...

Страница 38: ...generate output trigger signal when it sweeps One way The T200S sweeps from the start to the stop specified wavelength frequency at the specified sweep speed and returns to the initial wavelength at...

Страница 39: ...he Laser Output on page 39 5 If you want to arm the laser to the start position before sweeping touch the button The laser moves to the start position and is ready to perform a sweep sequence 6 Touch...

Страница 40: ...enu appears 3 In the Sweep Control area select Trigger In The icon appears in the Parameters area of the SWEEP window 4 Touch anywhere outside the menu to exit 5 Enable the laser output see Enabling D...

Страница 41: ...e time set in the Pause parameter moves back to the start position and waits for the next trigger to start the following sweep When it has completed the number of cycles the sweep stops and if the Aut...

Страница 42: ...WEEP mode window touch the button The mode setting menu appears 2 In the Sweep Control area select the Trigger Out check box The icon appears in the Parameters area of the SWEEP window to indicate tha...

Страница 43: ...internal referencing at any time as explained in the following procedure If the button appears on screen it means that the T200S needs internal referencing the T200S is fully usable but performances a...

Страница 44: ...instrument to which you want to output the signal meets the electrical requirements detailed for out BNC connector in Hardware Specifications on page 1 2 Connect the external instrument to the out BN...

Страница 45: ...turned off you cannot move the laser or enter the wanted power value The entered wavelength frequency is only applied if the laser is enabled To enable the laser output Touch the Enable button The la...

Страница 46: ...n the Settings window as explained in the following procedure To set the power unit 1 In the T200S task bar touch the button to display the Settings window and select the Instrument tab The Units area...

Страница 47: ...the signal meets the electrical requirements detailed for Power out BNC connector in Technical Specifications on page 1 2 Connect the external instrument to the Power out BNC connector located on the...

Страница 48: ......

Страница 49: ...the connection does not work contact your company network administrator To directly control the T200S from a computer 1 In the T200S task bar touch the button to display the Settings window and select...

Страница 50: ...it receives a command from an external controller When the remote mode is activated you cannot use the T200S GUI to control the T200S In remote mode the T200S GUI displays the Local mode button on the...

Страница 51: ...h the IEEE 488 2 standard These registers allow the controller to monitor events and get useful information on the status of the devices it controls Standard Event Status Register ESR Standard Event S...

Страница 52: ...ble shows the content of the Standard Event Status Enable Register ESE and Standard Event Status Register ESR Bit Weight Meaning 7 PON 128 Power ON Enable 6 URQ 64 User ReQuest Enable 5 CMD 32 CoMmanD...

Страница 53: ...Bit Weight Meaning 7 N U 128 Not used 6 N U 64 Not used 5 ESB 32 Event Summary Bit Enable 4 MAV 16 Message AVailable Enable 3 N U 8 Not used 2 EAV 4 Error event AVailable Enable 1 N U 2 Not used 0 N...

Страница 54: ...ument is warming up 10 1024 Always zero 9 512 INITializing the instrument is currently performing initialization 8 256 REFerencing the instrument is currently performing a reference 7 128 CORRecting t...

Страница 55: ...is not operational 9 512 Always zero 8 256 CALibration the calibration accuracy of the signal is of questionable quality 7 128 MODulation the modulation accuracy of the signal is of questionable quali...

Страница 56: ...type OUTP SYNC STAT ON The following table shows elements that are commonly used in the commands or queries syntax Item Meaning Enclose optional keywords or parameters Do not include square brackets i...

Страница 57: ...t with the same character set and to use supported characters while giving names to files or to other items Command Types The T200S uses the following types of commands and queries Sequential these co...

Страница 58: ...tion Error messages ending in a negative number are SCPI based errors Name Value Mnemonic Exa 1E18 EX Peta 1E15 PE Tera 1E12 T Giga 1E9 G Mega 1E6 MA Kilo 1E3 K Milli 1E 3 M Micro 1E 6 U Nano 1E 9 N P...

Страница 59: ...completely Cleaning the T200S Cleaning the Cover of the T200S If the external cover of the T200S becomes dirty or dusty clean it by following the instructions below Material needed Cleaning cloth Isop...

Страница 60: ...s of optical power you must regularly verify that the optical connector of the T200S is clean Handle optical fiber with appropriate care and preserve the integrity of optical connectors by keeping the...

Страница 61: ...ons 4a Hold the can of compressed air upright and spray the can into the air to purge any propellant 4b Spray the clean compressed air on the connector to remove any loose particles or moisture 5 Clea...

Страница 62: ...rts 4 Turn off the T200S see Turning off the T200S on page 18 5 On the front panel press the button to turn on the T200S The T200S automatically detects the pkg file on the USB device and starts the u...

Страница 63: ...system button The T200S verifies if your system is up to date This operation might take time If after 15 minutes the T200S is still checking for updates a problem may have occurred during the update c...

Страница 64: ...flat head screwdriver 1 or 2 replacement fuses for fuse type see Technical Specifications on page 3 To replace a fuse 1 Turn off the unit and unplug the power cord 2 Using a flat head screwdriver as a...

Страница 65: ...on All of these elements must be taken into consideration when determining the appropriate calibration interval of this particular EXFO unit Under normal use the recommended interval for your T200S is...

Страница 66: ......

Страница 67: ...tart the T200S as explained in Turning on the T200S and Accessing the GUI on page 16 Abruptly Turning off the T200S Emergency Shutdown In case of emergency presence of smoke flame or any immediate haz...

Страница 68: ...s to turn off the T200S 1b Press the On Off button to turn on the T200S 1c On the first sign that the T200S has started hold down the On Off button for 10 seconds to turn off the T200S 1d Press the On...

Страница 69: ...h Next 6 Read the warning and click Start to restore the unit to its initial state When the operation is complete the application prompts you 7 Touch OK The unit restarts If an error is display at sta...

Страница 70: ...og button The list of event log is displayed 3 Connect a USB storage device to one of the available USB ports 4 Touch the Save event log button The event log is saved at the root of the USB storage de...

Страница 71: ...XFO website www EXFO com en resources Contacting the Technical Support Group To obtain after sales service or technical support for this product contact EXFO at one of the following numbers The Techni...

Страница 72: ...ossibility of damage Pack the unit in its original packing material when shipping Avoid high humidity or large temperature fluctuations Keep the unit out of direct sunlight Avoid unnecessary shocks an...

Страница 73: ...ficial unauthorized or unintended by the original manufacturer Intermediaries using such channels to distribute products are considered to be part of the gray market hereafter unauthorized intermediar...

Страница 74: ...anges in the design or construction of any of its products at any time without incurring obligation to make any changes whatsoever on units purchased Accessories including but not limited to fuses pil...

Страница 75: ...e conditions under which it was observed 5 Return the equipment prepaid to the address given to you by support personnel Be sure to write the RMA number on the shipping slip EXFO will refuse and retur...

Страница 76: ...ervices exfo service centers EXFO Headquarters Service Center 400 Godin Avenue Quebec Quebec G1M 2K2 CANADA 1 866 683 0155 USA and Canada Tel 1 418 683 5498 Fax 1 418 683 9224 support exfo com EXFO Eu...

Страница 77: ...ned on the following pages Command Function Section CLS Clear status command CLS on page 72 ESE Standard event status enable command ESE on page 72 ESE Standard event status enable query ESE on page 7...

Страница 78: ...for the bits to be enabled in the Standard Event Status Register Type Overlapping Syntax ESE wsp register value Parameter s register value The register value represents the bit values of the Standard...

Страница 79: ...46 Type Overlapping Syntax ESE Parameter s None Response Syntax register value Response s register value The register value value represents the bit values of the Standard Event Status Enable register...

Страница 80: ...dard Event Status Enable Register ESE and Standard Event Status Register ESR on page 46 Example s ESR returns 33 the CME and OPC bits are set See Also ESE on page 73 ESR on page 74 IDN Description The...

Страница 81: ...Syntax OPC Parameter s None See Also OPC on page 75 WAI on page 79 OPC Description The OPC query makes possible the synchronization between the instrument and an external controller by reading the Ou...

Страница 82: ...ategy The RST command does the following Sets the device specific functions to a known state that is independent of the past use history of the device Forces the device into OCIS state Operation compl...

Страница 83: ...n page 47 Example s SRE 52 where 52 sets the ESB MAV and EAV bits See Also SRE on page 77 STB on page 78 SRE Description The SRE query returns the current content of the Service Request Enable Registe...

Страница 84: ...ns 68 the MSS and EAV bits are set See Also SRE on page 77 TRG Description This command causes the immediate exit from the event detector of a trigger sequence if the system is waiting for the BUS eve...

Страница 85: ...al Commands WAI Description The WAI command prevents the device from executing any further commands or queries until the no operation pending flag becomes TRUE Type Overlapping no query Syntax WAI Par...

Страница 86: ...on page 90 OUTPut OUTPut Commands and Queries on page 92 SOURCE SOURce Commands and Queries on page 96 STATus STATus Commands and Queries on page 126 SYSTem SYSTem Commands and Queries on page 128 TRI...

Страница 87: ...Yer SOURce wsp source Parameter s source Valid arm event source The allowed values are BUS the sweep sequence is started by the TRG command Once initiated with the INITiate IMMediate ALL on page 89 th...

Страница 88: ...source BUS the sweep sequence is started by the TRG command see TRG on page 78 Once initiated with the INITiate IMMediate ALL on page 89 the instrument waits for the TRG command to start the sequence...

Страница 89: ...5 DATA TIME see p 86 WARMup ABORt see p 86 CALibration ALL Description This command starts a full wavelength referencing of the instrument During the execution of this command bit 0 CALibrating is set...

Страница 90: ...None Response Syntax temperature Response s temperature Floating point number corresponding to the temperature value measured during wavelength referencing The unit of the returned value depends on t...

Страница 91: ...integer corresponding to the year of the current calibration month Integer corresponding to the month of the current calibration in the range 01 to 12 01 corresponds to January 02 to February and so o...

Страница 92: ...r corresponding to the number of seconds in the range 0 to 59 Example s CAL DATA TIME returns 14 57 09 See Also CALibration ALL on page 83 CALibration DATA DATE on page 85 CALibration DATA TEMPerature...

Страница 93: ...pe Format of data readings Possibles values are ASCii numeric data is formatted as ASCII bytes such as NR1 NUMERIC RESPONSE DATA NR2 NUMERIC RESPONSE DATA or NR3 NUMERIC RESPONSE DATA representation a...

Страница 94: ...NUMERIC RESPONSE DATA or NR3 NUMERIC RESPONSE DATA representation as approriate PACKed data is formatted as a series of 8 bit binary data bytes length If the data is returned in ASCii format the lengt...

Страница 95: ...Tiate IMMediate ALL Description This command completes one full trigger system cycle and returns to IDLE on completion This command initiates a sweep sequence This command is an event and has no assoc...

Страница 96: ...uery returns the power read at the source output This query has no associated RST condition Syntax MEASure SCALar POWer DC Parameter s None Response Syntax power Response s power Floating point number...

Страница 97: ...ASure SCALar TEMPerature DC Parameter s None Response Syntax temperature Response s temperature Floating point number corresponding to the temperature measured in the instrument The unit of the return...

Страница 98: ...laser output At RST this value is set to ON During the execution of this command bit 1 SETTling is set in the Operation Status Condition register see Operational Questionable Status Reporting on page...

Страница 99: ...ration on the specified monitoring BNC connector located on the rear panel MONitor1 is the out BNC connector for spectrum monitoring MONitor2 is the Power out BNC connector for power monitoring During...

Страница 100: ...oring MONitor2 is the Power out BNC connector for power monitoring At RST this value is set to OFF Syntax OUTPut MONitor1 2 STATe Parameter s None Response Syntax state Response s state State of the s...

Страница 101: ...tput is open Example s OUTP SYNC STAT OFF closed See Also OUTPut MONitor1 2 STATe on page 93 OUTPut SYNChronization STATe on page 95 OUTPut SYNChronization STATe Description This query indicates if th...

Страница 102: ...EFault see p 105 OPTimization BCOMpensation STATe state see p 106 STATe see p 106 OLINewidth STATe state see p 107 STATe see p 107 SWEep CYCLe CONTinuous state see p 108 CONTinuous see p 108 COUNt cou...

Страница 103: ...pecific Commands SPEed LIST see p 121 STARt start unit MAXimum MINimum DEFault see p 122 STARt MAXimum MINimum DEFault see p 123 STOP stop unit MAXimum MINimum DEFault see p 124 STOP MAXimum MINimum D...

Страница 104: ...ible frequency value depending on the laser model MINimum Sets the emission frequency to the minimum possible frequency value depending on the laser model DEFault Sets the emission frequency to the de...

Страница 105: ...for the laser model MINimum The query returns the minimum possible frequency value for the laser model DEFault The query returns the default frequency value for the laser model Response Syntax frequen...

Страница 106: ...OURce FREQuency CW FIXed STEP WIDTh on page 101 MIN and MAX parameters unit Unit of the frequency step HZ preceded by a possible multiplier Possible multipliers are listed in Unit Multipliers on page...

Страница 107: ...s the maximum possible frequency value for the laser model MINimum The query returns the minimum possible frequency value for the laser model DEFault The query returns the default frequency value for...

Страница 108: ...ers unit Unit of the start frequency HZ preceded by a possible multiplier Possible multipliers are listed in Unit Multipliers on page 52 MAXimum Sets the start frequency to the maximum possible value...

Страница 109: ...l MINimum The query returns the minimum possible start frequency for the laser model DEFault The query returns the default start frequency for the laser model Response Syntax start Response s start Fl...

Страница 110: ...ameters unit Unit of the stop frequency HZ preceded by a possible multiplier Possible multipliers are listed in Unit Multipliers on page 52 MAXimum Sets the stop frequency to the maximum possible valu...

Страница 111: ...del MINimum The query returns the minimum possible stop frequency for the laser model DEFault The query returns the default stop frequency for the laser model Response Syntax stop Response s stop Floa...

Страница 112: ...pensation enabled See Also SOURce OPTimization BCOMpensation STATe on page 106 SOURce OPTimization OLINewidth STATe on page 107 SOURce OPTimization BCOMpensation STATe Description This query returns t...

Страница 113: ...Rce FREQuency CW FIXed on page 98 SOURce OPTimization BCOMpensation STATe on page 106 SOURce OPTimization OLINewidth STATe on page 107 SOURce WAVelength CW FIXed on page 115 SOURce OPTimization OLINew...

Страница 114: ...E CYCL CONT ON SOUR SWE CYCL CONT returns 1 continuous cycle enabled See Also SOURce SWEep CYCLe CONTinuous on page 108 SOURce SWEep CYCLe COUNt on page 109 SOURce SWEep CYCLe CONTinuous Description T...

Страница 115: ...nt Number of sweep cycles you want the instrument to perform To know the valid count range use the query SOURce SWEep CYCLe COUNt on page 110 MIN and MAX parameters MAXimum Sets the number of sweep cy...

Страница 116: ...ns the maximum possible number of sweep cycles MINimum The query returns the minimum possible number of sweep cycles DEFault The query returns the default number of sweep cycles Response Syntax count...

Страница 117: ...now the valid time range use the query SOURce SWEep HOLDoff on page 112 MIN and MAX parameters unit Unit of the pause value S preceded by a possible multiplier Possible multipliers are listed in Unit...

Страница 118: ...Parameter s MAXimum The query returns the maximum possible pause duration in second MINimum The query returns the minimum possible pause duration in second DEFault The query returns the default pause...

Страница 119: ...s SOURce FREQuency SWEep STARt on page 102 and SOURce WAVelength SWEep STARt on page 122 STOP on sweep sequence completion the source stays at the sweep stop position If the sweep sequence is complete...

Страница 120: ...SOURce FREQuency SWEep STARt on page 102 and SOURce WAVelength SWEep STARt on page 122 STOP on sweep sequence completion the source stays at the sweep stop position If the sweep sequence is completed...

Страница 121: ...le wavelength value depending on the laser model MINimum Sets the emission wavelength to the minimum possible wavelength value depending on the laser model DEFault Sets the emission wavelength to the...

Страница 122: ...for the laser model MINimum The query returns the minimum possible wavelength value for the laser model DEFault The query returns the default wavelength value for the laser model Response Syntax wave...

Страница 123: ...query SOURce WAVelength CW FIXed STEP WIDTh on page 118 MIN and MAX parameters unit Unit of the wavelength step M preceded by a possible multiplier Possible multipliers are listed in Unit Multipliers...

Страница 124: ...imum possible wavelength value for the laser model MINimum The query returns the minimum possible wavelength value for the laser model DEFault The query returns the default wavelength value for the la...

Страница 125: ...econd To know the valid speed range use the query SOURce WAVelength SWEep SPEed on page 120 MIN and MAX parameters MAXimum Sets the speed to the maximum possible value MINimum Sets the speed to the mi...

Страница 126: ...ximum possible speed in meter per second MINimum The query returns the minimum possible speed in meter per second DEFault The query returns the default speed in meter per second Response Syntax speed...

Страница 127: ...eter s None Response Syntax speed Response s speed Value corresponding to the vector of speed values in meter per second You can use the FORMat DATA PACK command to set list data transfer in compresse...

Страница 128: ...rameters unit Unit of the start wavelength M preceded by a possible multiplier Possible multipliers are listed in Unit Multipliers on page 52 MAXimum Sets the start wavelength to the maximum possible...

Страница 129: ...MINimum The query returns the minimum possible start wavelength for the laser model DEFault The query returns the default start wavelength for the laser model Response Syntax start Response s start F...

Страница 130: ...meters unit Unit of the stop wavelength M preceded by a possible multiplier Possible multipliers are listed in Unit Multipliers on page 52 MAXimum Sets the stop wavelength to the maximum possible valu...

Страница 131: ...el MINimum The query returns the minimum possible stop wavelength for the laser model DEFault The query returns the default stop wavelength for the laser model Response Syntax stop Response s stop Flo...

Страница 132: ...ondition register associated with the operation status structure Reading the condition register is nondestructive Type Overlapping query only Syntax STATus OPERation CONDition Parameter s None Respons...

Страница 133: ...condition register is nondestructive Type Overlapping query only Syntax STATus QUEStionable CONDition Parameter s None Response Syntax condition Response s condition Unique integer in the range 0 to 6...

Страница 134: ...ted RST condition Syntax SYSTem DATE wsp year month day Parameter s year Four digit number corresponding to the year in the instrument s date month Number corresponding to the month in the instrument...

Страница 135: ...ear in the instrument s date month Integer corresponding to the month in the instrument s date in the range 1to 12 1 corresponds to January 2 to February and so on day Integer corresponding to the day...

Страница 136: ...instrument specific The zero value indicates that no error or event has occurred description Quoted string containing a description followed by an optional information text Each code has a unique and...

Страница 137: ...ST TIME 16 58 41 See Also SYSTem DATE on page 128 SYSTem TIME on page 131 SYSTem TIME Description This query returns the current time from the instrument s internal clock This query has no associated...

Страница 138: ...e s version Fixed point number corresponding to the year followed by the revision number of the SCPI standard to which the instrument complies The year represents the year version and the revision rep...

Страница 139: ...tax TRIGger SEQuence SOURce wsp source Parameter s source Valid trigger event source The allowed values are BUS the source signal comes from the command TRG on page 78 EXTernal the source signal comes...

Страница 140: ...ponse s source Selected trigger event source BUS the source signal is generated by the command TRG on page 78 EXTernal the source signal comes from the external signal jack via the Trig In port locate...

Страница 141: ...unit see p 136 TEMPerature see p 137 UNIT POWer Description This command selects a default unit for commands that set absolute power At RST this value is set to DBM Syntax UNIT POWer wsp unit Paramet...

Страница 142: ...to watt Example s UNIT POW DBM UNIT POW returns DBM See Also UNIT POWer on page 135 UNIT TEMPerature on page 137 UNIT TEMPerature Description This command selects a default unit for commands that set...

Страница 143: ...this value is set to CEL Syntax UNIT TEMPerature Parameter s None Response Syntax unit Response s unit Default unit for commands that set temperature C CEL the temperature unit is set to Celsius F FAR...

Страница 144: ......

Страница 145: ...ned in at least one of the homogeneous materials used for this part is above the limit requirement of GB T 26572 Due to the limitations in current technologies parts with the X mark cannot eliminate h...

Страница 146: ...EXFO com info exfo com CORPORATE HEADQUARTERS 400 Godin Avenue Quebec Quebec G1M 2K2 CANADA Tel 1 418 683 0211 Fax 1 418 683 2170 TOLL FREE USA and Canada 1 800 663 3936 2022 EXFO Inc All rights reser...

Отзывы: