SCPI Command Reference
104
T200S
T200S Specific Commands
[SOURce]:FREQuency:SWEep:STOP
Description
This command sets the stop frequency of a swept signal.
At *RST, this value will be set to MAXimum.
Syntax
[:SOURce]:FREQuency:SWEep:STOP<wsp><stop>[<unit>]|
MAXimum|MINimum|DEFault
Parameter(s)
³
stop
:
Value corresponding to the stop frequency in Hz (if no unit is
specified).
To know the valid frequency range, you can use the query
[SOURce]:FREQuency:SWEep:STOP?
MAX parameters).
³
unit
:
Unit of the stop frequency: HZ preceded by a possible
multiplier. Possible multipliers are listed in
³
MAXimum
Sets the stop frequency to the maximum possible value.
³
MINimum
Sets the stop frequency to the minimum possible value.
³
DEFault
Sets the stop frequency to the default value.
To know the default value, you can use the query
[SOURce]:FREQuency:SWEep:STOP?
parameter).
Example(s)
SOUR:FREQ:SWE:STOP 197.5 THZ
SOUR:FREQ:SWE:STOP? returns 1.014
See Also
[SOURce]:FREQuency:SWEep:STARt
[SOURce]:FREQuency:SWEep:STOP?