Remotely Controlling the T200S
52
T200S
Communication Principle
Unit Multipliers
The following table illustrates the possible unit multipliers:
Error Messages
System and device-specific errors are managed by your unit. The generic format for error
messages is illustrated in the following figure.
As shown in the above figure, the message contains three parts:
³
error number
³
error description
³
device-dependent information
Error messages ending in a negative number are SCPI-based errors.
Name
Value
Mnemonic
Exa
1E18
EX
Peta
1E15
PE
Tera
1E12
T
Giga
1E9
G
Mega
1E6
MA
Kilo
1E3
K
Milli
1E-3
M
Micro
1E-6
U
Nano
1E-9
N
Pico
1E-12
P
Femto
1E-15
F
Atto
1E-18
A
I
MPORTANT
If an error occurs in a command of a program message (several commands
separated by ";"), the program does not stop: all the commands of the program
message are executed. As a consequence, several errors may be present in the error
queue. You need to check the error bit to make sure that the error queue is empty.
<Error
number>
,
"
<Error
description>
;
"
<Device
dependent
information>