Remotely Controlling the T100S-HP
T100S-HP Programming Guide
11
1.2.2
Setting the GPIB Address
Subject
The default GPIB address of the T100S-HP is factory-set to 10. This section explains how
to modify it (possible values are 1 to 30).
Once changed, the new GPIB address is stored in flash memory, and becomes the new
default system setting. The T100S-HP retains the new GBIP address even after system
turn-off.
To void address conflicts, you must make sure that your T100S-HP GPIB address is
different from the address of any other device already connected to the GPIB port.
Procedure
Setting the GPIB Address from the Front Panel
1.
Press the
2nd
+
Remote
keys.
The current GPIB address is displayed as the display prompts you to enter the new
value at the numeric keypad.
2.
Enter the new GPIB address, in the range of 1 to 30.
3.
Press
Enter
to confirm the change.
The
DATA
area display prompts you to activate/deactivate the RS-232C remote
mode.
4.
Press
Esc
.
Setting the GPIB Address in Remote Mode
Use the command:
GPAD=xx
(see section
1.2.3
Connecting the T100S-HP to an IEEE-488.1 Controller
Subject
The GPIB port enables you to connect the T100S-HP to a computer to control it via
remote commands.
Before Starting
Make sure you have a GPIB cable to link the T100S-HP to an IEEE-488.1 controller (GPIB
PCI card or GPIB-USB-HS from National Instrument) connected to your computer.
Procedure
1.
Connect the GPIB port of the T100S-HP to the IEEE-488.1 controller connected to
your computer via the GPIB cable.
2.
Use the authorized remote GPIB commands detailed in the present guide to remotely
control the T100S-HP.