Test Results
Power Blazer
381
Alarms/Errors Overview
ODUx-TCM
Alarms
BDI
(Backward Defect Indication)
RX:
The BDI bit in the TCM overhead field Byte 3, bit 5 is “1” for at least
5 consecutive frames.
TX
: Generates a "1" in the BDI bit of the TCM overhead field (byte 3,
bit 5) continuously.
BIAE
(Backward Incoming Alignment Error)
RX:
The BEI/BIAE bits in the TCM overhead field Byte 3, bits 1 to 4 are
“1011” for at least 3 consecutive frames.
TX
: Generates "1011" in the BEI/BIAE bits of the TCM overhead (byte 3,
bits 1 to 4) continuously.
IAE
(Incoming Alignment Error)
RX:
The STAT information in the TCM is “010” for at least 3 consecutive
frames.
TX
: Generates "1" in the IAE bit of the TCM overhead (byte 3, bit 6)
continuously.
LTC
(Loss of Tandem Connection)
RX:
The STAT information in the TCM Byte 3, bits 6, 7, and 8 are “000”
for at least 3 consecutive frames.
TX
: Generates "000" in the STAT field of TCM overhead (byte 3, bits
6 to 8) continuously.
TIM
(Trace Identification Mismatch)
RX
: The SAPI and/or DAPI do not math the expected SAPI and/or DAPI.
This alarm is only available when the Enable TIM SAPI and/or DAPI
check boxes are selected from
Содержание Power Blazer 88000 Series
Страница 1: ...User Guide 88000 Series Power Blazer HIGH SPEED MULTISERVICE TEST MODULE ...
Страница 12: ......
Страница 16: ......
Страница 22: ......
Страница 28: ......
Страница 38: ......
Страница 58: ...Test Setup Test Applications 46 88000 Series OTN BERT ...
Страница 64: ...Test Setup Test Applications 52 88000 Series SONET SDH BERT For SDH BERT ...
Страница 69: ...Test Setup Test Applications Power Blazer 57 SONET SDH DSn PDH BERT For SDH ...
Страница 550: ......
Страница 670: ......
Страница 674: ......
Страница 686: ......
Страница 707: ...Glossary Power Blazer 695 10G Ethernet Client The following figure presents a typical network application ...
Страница 721: ...Glossary Power Blazer 709 Ethernet Cables Straight Through Cable 1000 Mbit s Crossover Cable 1000 Mbit s ...
Страница 765: ...Glossary Power Blazer 753 SONET DSn SDH PDH The GUI Grid indicates the TUG 2 x and TUG 3 x values ...