Introducing the IQS-9100 Optical Switch
12
IQS-9100
Typical Applications
Measuring Optical Switch Port Insertion Loss
The insertion loss (IL) of any switch port can be measured using a light
source, power meter, and the IQS-9100 Optical Switch.
To measure optical switch port insertion loss:
1.
Connect the light source output port to the power meter detector port
using two test jumpers linked with a bulkhead connector.
2.
Store the power reading obtained as a reference value in the power
meter.
Light Source
IQS-2100
ACTIVE
Power Meter
IQS-1100
Bulkhead connector
Light source
output port
Power meter
detector port