Using IQS Products in an Automated Test Environment
178
IQS-600
Managing the States of Your IQS-600 Integrated Qualification System
The following diagram shows the possible transitions between the
different states.
No.
Possible in GPIB if:
Possible in ActiveX by setting:
1
The remote interface message (REM) is set
to True.
The device is addressed to listen (MLA).
The return to local interface message (rtl) is
set to False.
The
RemoteState
property to
True.
2
The remote interface message (REM bus
line) is set to True.
The local lockout interface message (LLO)
is set to True.
The
LockoutState
property to
True.
3
One of the following conditions is true:
The go to local interface message (GTL) is
received and the device is addressed to
listen (MLA).
The Clear Lockout/Set Local message is
received (the REM bus line is set to False).
The return to local interface message (rtl) is
set to True and the local lockout interface
message (LLO) is set to False.
The
RemoteState
property to
False.
Local
Local Lockout
Remote Lockout
Remote
3
5
2
9
6
8
1
7
4
Содержание IQS-600 Series
Страница 1: ...User Guide Integrated Qualification System IQS 600...
Страница 22: ......
Страница 28: ......
Страница 106: ......
Страница 132: ......
Страница 194: ...Using IQS Products in an Automated Test Environment 186 IQS 600 Standard Status Data Structure...
Страница 252: ......
Страница 274: ......
Страница 336: ......
Страница 366: ...Using the IQS 9600 Utility Module 358 IQS 600...
Страница 367: ...Using the IQS 9600 Utility Module Integrated Qualification System 359...
Страница 368: ......