Performing Acquisitions
60
IQS-1700
Selecting the Sampling Type
The acquisition rate applies to all channels when using a multichannel
power meter.
You can also change the continuous sampling rate in the
Instrument
function tab in the same manner.
If you select
Single
, select the rate to use with the arrow buttons.
This rate is a divider of the maximum rate (5208 Hz). For example,
5208 Hz has a divider of 1, 2604 has a divider of 2 and so forth. To add
or remove dividers from the list of available values, see
on page 61. The acquisition rate applies to all
channels when using a multichannel power meter.
If you select
User Defined
, select the rate by entering a value in the
1/
list, then select a time unit using the arrow buttons. Remember to click
to confirm your settings.
Note:
Use a lower sampling rate with averaging set to active (see
on page 51) for greater repeatability when measuring very
low power.