When the averaging function is enabled on the IQS-1700 High Performance
Power Meter, the most recent measurement samples, for which you can
set the number, are used to compute an unweighted average. This average
is displayed as the measured value. The
AVG
marker will also be displayed
to the right of the value for the matching channel, indicating that averaging
is enabled.
Averaging can be done on a minimum of 2 and a maximum of 1000 points
(you can enter 1 as a value, but the averaging will be performed only on
that value and will be rather useless). The calculation uses watt as unit, but
the result will be converted into the unit you have selected for your
acquisition (W, W/W, dBm or dB) afterwards.
The formulas used to calculate measurement averaging depend on the
number of the current acquisition.
If n
M, then
Where
n
is the number of the current acquisition
M
is the number of points to average (value you set)
x[n]
is the current sample in watts
y[n]
is the result of the present average in watts
y[n-1]
is the value before the average in watts
y n
1
M
----- x n
1
1
M
-----
–
y n 1
–
+
=
Содержание IQS-1700
Страница 1: ...User Guide High Performance Power Meter and Optical Head for IQS 600 IQS 1700...
Страница 5: ...Contents High Performance Power Meter v Index 189...
Страница 8: ......
Страница 14: ......
Страница 16: ......
Страница 44: ......
Страница 64: ......
Страница 80: ...Performing Acquisitions 72 IQS 1700 Consulting Acquired Data To view your acquisition data select the Data tab...
Страница 96: ......
Страница 114: ......
Страница 196: ......