Remote Control
Tunable Laser Source
33
Specific Commands
Specific Commands
CALibration:WAVElength
Description
This command introduces an offset into the IQ-2600B to correct
a wavelength deviation as measured by a wavelength meter.
Syntax
CAL:WAVE<space><wavelength>
Parameters
The <wavelength> parameter is a value indicating the
wavelength in nm. The format is 9999.999.
Example
CAL:WAVE 1530.000
Note
This command does not support THz as units.
OUTPut:ASE?
Description
This query returns a value indicating whether the ASE option is
available on the tunable laser source.
Syntax
OUT:ASE?
Response
A boolean value indicating whether the ASE option is available
on the IQ-2600B Tunable Laser Source:
“0”-the ASE option is not available
“1”-the ASE option is available
Note
Since the ASE mode is not available with the IQ-2600B, the
return value will always be “0”.
See also
OUTP:MODE, OUTP:MODE?, OUTP:STAT and OUTP:STAT?
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com