![EXFO FTV-700 series Скачать руководство пользователя страница 59](http://html1.mh-extra.com/html/exfo/ftv-700-series/ftv-700-series_user-manual_2450048059.webp)
Testing Fibers in Advanced Mode
OTDR
49
Setting Pass/Fail Thresholds
To set pass/fail thresholds:
1.
From the button bar, select
OTDR Setup
, then select the
Event Table
tab.
2.
From the
Wavelength
list, select the wavelength for which you want to
set thresholds.
3.
Select the boxes corresponding to the thresholds that you want to use,
and enter the desired values in the appropriate fields.
Note:
If you no longer want the application to take into account a particular
threshold, simply clear the corresponding box.
4.
If you want the application to display messages when events fail the
test, select
Display Pass/Fail message
.
Threshold to set
Value associated
with the
threshold to set