Chromatic Dispersion Analyzer
87
A Technical Specifications
I
MPORTANT
The following technical specifications can change without notice.
The information presented in this section is provided as a reference
only. To obtain this product’s most recent technical specifications,
visit the EXFO Web site at
www.exfo.com
.
SPECIFICATIONS
a
Model
FTB-5800
Wavelength range (nm)
1530 to 1625
1200 to 1700
b
Wavelength step (nm)
Minimum
0.1
Measurement points
Maximum
950, user-definable
Dynamic range
c
(dB)
42
Wavelength uncertainty
d
(accuracy) (nm)
0.1
Dispersion uncertainty
d
(accuracy) (ps/nm)
20 km of G.652
1.6
120 km of G.652
3.1
20 km of G.655
1.9 (guaranteed)
20 km
80 km
120 km
Dispersion repeatability
d
(ps/nm)
0.04
0.2
1.1
Zero-dispersion wavelength
λ
0
repeatability
d
(nm)
0.1
0.14
0.8
Dispersion slope repeatability
λ
0
d
(%)
0.03
0.05
0.25
Minimum fiber length (km)
< 1
Maximum fiber length (km)
e
> 5400
Measurement time per point
e
(s)
Minimum
< 1
NOTES
a. All specifications are typical with 4 seconds averaging time per point (where applicable),
at a temperature of 23 ºC ± 1 ºC, with FC connectors and after warmup time.
b. Displayed range. Values may be extrapolated.
c. Dynamic range is defined as the difference between the strongest signal and the
weakest signal the receiver can detect. Extra averaging may be required.
Uncertainty (accuracy) is not guaranteed at limits of range.
d. C+L band.
e. Including EDFAs.
f. Additional gain setting time may be required prior to the first point of each band.
GENERAL SPECIFICATIONS
Size (H x W x D) (module)
9.6 cm x 10 cm x 26 cm
(3
3
/
4
in x 3
15
/
16
in x 10
1
/
4
in)
Weight (module)
2 kg
(4.5 lb)