CT440
83
9
Troubleshooting
Warning & Error Messages
The following table lists all the possible warning and error messages, and how to handle
them.
Trouble
Possible Resolution
The CT440 is not recognized by the
computer
See
The CT440 GUI does not start
See
The CT440 comes offline during
use
Restart both GUI and instrument to detect the CT440 again.
The GUI displays an error or
warning message
See
The
icon appear in the GUI (at
the left of the date and time)
See
Error or warning
Possible Cause
Possible solution
The
icon appear in the
GUI, at the left of the date
and time (on CT440 with
PDL option only).
Some points are missing in
the PDL calculation: the
CT440-PDL could not
calculate PDL for a few
measurement points.
Verify that the CT440-PDL is set on a flat
stable surface without vibration.
Verify that optical connectors are clean.
Verify that optical connectors are properly
connected to the CT440-PDL and tightly
locked in position.
Use the 6-state PDL measurement
method: in the
Scan Parameters
window,
select
6 States
in the
PDL Method
list (see
No Supported Languages
- Unable to open resource
files
The CT440 GUI uses
LabVIEW2013 RunTime. If
the CT440 GUI does not
start, it may be necessary to
repair the RunTime files.
From the Windows Control panel, in the
Remove Programs
window, right-click
the National Instruments program and
select
Unistall/Change
.
Repair the NI LabVIEW Runtime Engine
2013.
Restart the PC.
[warning code 100]
Mode hops on the scan
Mode hopping occurred
during the scan on one TLS.
Make sure the source performances meet the
[warning code 101]
Scan speed too low
The mean sweeping
speed of the TLS is too
low (< 8nm/s)
The reference file is not
located in the correct
folder.
Raise the scan speed.
Make sure the reference file is located in
the proper
Calib
folder:
C:\Users\Public\Documents\EXFO\CT440\
Calib