Configuration Functions
CT440_SetExternalSynchronizationIN
Applicability
All models of CT440.
Description
This function configures the CT440 to run a scan triggered by the
connected tunable laser (
Trigger In
BNC connector located on the
rear panel of the instrument).
For more details, see
In the GUI, this function corresponds to the
Single Trig
or
Continuous Trig
scanning modes in the
Scan Parameters
window.
On the T100S-HP, the
BNC output must be activated as a trigger (for
more details, see
T100S-HP User Manual
).
Declaration
int32_t CT440_SetExternalSynchronizationIN
(uint64_t uiHandle,
rEnable eEnable);
Parameters
uiHandle
Handle returned from
CT440_Init
Type: Input
Data type: unsigned int64
eEnable
Enables/Disables the laser output.
Possible values: see
Type: Input
Data type: rEnable
If Enable, the CT440 waits for an external trigger on the
Trigger In
BNC port to start measurements.
Return value
0: the operation succeeded.
-1: the operation failed.
Data type: int32
Example
CT440_SetExternalSynchronizationIN(uiHandle,ENABLE);