background image

8

Exergen TAT-5000S-RS232

Q: How does the temperature from a temporal 
scanner relate to core temperature?

A:

 Temporal artery temperature is considered 

a core temperature because it has been 
demonstrated as accurate as the temperature 
measured by a pulmonary artery and esophageal 
catheter, and as accurate as a rectal temperature 
on a stable patient. Rule of thumb: Rectal 
temperature is about 0.5°C (1°F) higher than an 
oral temperature and 1°C (2°F) higher than an 
axillary temperature. It will be easy to remember 
if you think of core temperature as a rectal 
temperature, and apply the same protocol you 
would use for a rectal temperature.

If your thermometer is marked Arterial/Oral 
and has a serial number beginning with “O” 
(standard model start with “A”), it is programmed 

to compute the normal average cooling effect 

at the mouth, and automatically reduces the 
higher arterial temperature by that amount. 
This calibration allows the hospital to maintain 
existing protocols for fever workups based 
on oral temperature, and results in a reading 
consistent with the 37°C (98.6°F) mean normal 
oral temperature, in the range of 35.9 - 37.5°C 
(96.6 - 99.5°F) you now see.

Q: What should I do if I get an abnormally 

high or low reading, how do I confirm my 

reading?

•  Repeat the reading with the same Temporal 

Scanner; a correct reading will be reproducible.

•  Repeat the reading with another Temporal 

Scanner. Two Temporal Scanners with the 

same reading will confirm the reading.

•  Sequential readings on the same patient in 

rapid succession will cool the skin; it is best to 
wait about 30 seconds for the skin to recover 
from the cold probe.

Possible causes of abnormal readings.

Type of 
Abnormal 
Temperature

Possible  
Cause

Helpful  
Hint

Abnormally  
Low Temperature

Dirty Lens

Clean lens of 
scanner every two 
weeks.

Releasing the 
button before 

finished measuring

Release the button 

after finished 

measuring.

Measuring when 
an ice pack or wet 
compress is on the 
forehead

Remove ice pack 
or wet compress, 
wait 2 minutes, 
and re-take 
temperature.

Measuring a 
completely 
diaphoretic patient

Complete 
diaphoresis 
includes 
diaphoresis of area 
behind the ear 
and suggests that 
the temperature 
is rapidly 
dropping.  Use an 
alternative method 
of temperature 
measurement 
in these cases 
until the patient 
is dry and the 
temporal artery 
measurement can 
be repeated.

Improperly 
scanning down the 
side of the face

Scan straight 
across forehead. 
The temporal 
artery is closest to 
skin in that area.

Abnormally  
High Temperature

Anything covering 
the area to be 
measured would 
insulate and 
prevent heat 
from dissipating, 
resulting in false 
high readings.

Confirm 

measurement site 
has not recently 
been in contact 
with heat insulators 
such as hats, 
blankets, and hair.
Scan the area not 
covered or wait 
about 30 seconds 
for the previously 
covered area to 
equilibrate to the 
environment.

Frequently Asked Questions

Содержание TemporalScanner TAT-5000S-RS232 Series

Страница 1: ...1 Exergen TAT 5000S RS232 TemporalScanner TAT 5000S RS232 Series Changing the Way the World Takes Temperature Accurate Temperature with a Gentle Forehead Scan Operation Manual ...

Страница 2: ...l thermal radiation emitted from the skin It has been clinically proven in premier university hospitals to be more accurate than ear thermometry and better tolerated than rectal thermometry and is supported by more than 70 peer reviewed published studies covering all ages from premature infants to geriatrics in all clinical care areas It is a superior method for patient and clinician alike A 40 pa...

Страница 3: ...temperature of patients Clinical environments include areas where medical professionals are providing medical services for patients including hospitals outpatient clinics primary care offices and other settings where temperature is taken as part of patient care Clinical environments include Emergency Medical Services environments Additionally the TAT 5000S series thermometers are not for use aboar...

Страница 4: ...scribed above see the FAQ section of this manual Additionally ensure that you are not in the presence of electromagnetic disturbances If you have any additional questions regarding use or care of the thermometer please see www exergen com or call customer service at 617 923 9900 SAVE THESE INSTRUCTIONS Use of this equipment adjacent to or stacked with other equipment other than TAT 5000S compatibl...

Страница 5: ...can for the highest temperature peak as long as the button is depressed Clicking Each fast click indicates a rise to a higher temperature similar to a radar detector Slow clicking indicates that the instrument is still scanning but not finding any higher temperature To Retain or Lock Reading The reading will remain on the display for 30 seconds after button is released If measuring room temperatur...

Страница 6: ...nt here is typically all that is required Measure straight across the forehead and not down side of face At mid line the temporal artery is about 2 mm below the surface but can go deeply below the surface on the side of the face If the temporal artery is covered then the area behind the ear if exposed can be an alternate site Brush the hair aside if covering the area to be measured Measurement sit...

Страница 7: ...uch behind ear halfway down the mastoid process and slide down to the soft depression behind the earlobe Think of a sweatband Measure straight across the forehead and not down the side of the face At mid line the temporal artery is about 2 mm below the surface but can go deeply below the surface on the side of the face Step 3 Release button remove from head and read Minimum measuring time 2 second...

Страница 8: ...ible Repeat the reading with another Temporal Scanner Two Temporal Scanners with the same reading will confirm the reading Sequential readings on the same patient in rapid succession will cool the skin it is best to wait about 30 seconds for the skin to recover from the cold probe Possible causes of abnormal readings Type of Abnormal Temperature Possible Cause Helpful Hint Abnormally Low Temperatu...

Страница 9: ... cotton swab dipped in alcohol in accordance with the instruction label on the instrument see below Use only light force for cleaning to avoid damaging the lens Water can be used to remove any residual film left by the alcohol Do not use bleach or other cleaning solutions on the sensor lens Sterilization Sterilization is not recommended for cabled versions of the TemporalScanner Calibration Factor...

Страница 10: ...ergen for repair if error message persists Scanning Normal Operation The TemporalScanner can be used in either F or C To convert from one scale to the other the only tools necessary are a paper clip and the tip of a small screwdriver For F C Conversion Insert the end of a bent paper clip into the pinhole on the side to release and remove the cover Remove the battery from the compartment Locate the...

Страница 11: ...ed separation distance d 1 2 P1 2 d 1 2 P1 2 80 MHz to 800MHz d 1 2 P1 2 800MHz to 2 7 GHz Where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m Field strength from the fixer RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level ...

Страница 12: ...cteris tic of a typical location in a typical health care environment Note UT is the a c mains voltage prior to the application of the test level Recommended separation distances between portable and mobile RF communication equipment and the TAT 5000S Series The TAT 5000S series forehead thermometer is intended for use in an electromagnetic environment in which radiated RF disturbances are control...

Страница 13: ...onse Time 0 04 seconds Time Displayed On Screen 30 seconds Size Instrument 20 cm X 4 6 cm X 4 cm 7 9 X 1 8 X 1 6 Cable 0 8 m 32 retracted Weight 0 3 kg 0 7 lb EMI and RFI Protection Stainless steel enclosure on upper part inside of casing Storage Conditions 20 to 50 C 4 to 122 F Display Type and Size Large bright LED s Construction Method Industrial duty impact resistant casing Chemically resistan...

Страница 14: ...lease the battery cover then move battery out of the way 2 Depress black release button and remove cable while holding the release button down 3 Orient cable plug properly for proper fit into the receptacle hole and install replacement cable 4 Replace the battery into the compartment Re install battery cover Exergen p n Description 124330 QR Replacement Cable Generic ...

Страница 15: ...tection should they be preferred for certain patient populations and are still very cost effective These options include resposable caps and full instrument sheaths the sheaths being mainly used for isolation patients Using the Resposable Disposable Caps 1 Apply cap by pushing onto the probe head with fingers 2 Remove cap by pushing edge forward with thumb 3 Caps may be reused on the same patient ...

Страница 16: ...lation Proof Type BF Applied Part Battery Operated MEDICAL ELECTRICAL EQUIPMENT ANSI AAMI ES60601 1 2005 R 2012 3rd Edition including Amendment 1 CAN CSA C22 2 No 60601 1 2014 IEC 60601 1 6 ISO 80601 2 56 Particular Requirements For Basic Safety and Essential Performance of Clinical Thermometers For Body Temperature Measurement The CLINICAL THERMOMETER is an ADJUSTED MODE CLINICAL THERMOMETER Corr...

Отзывы: