background image

7

The Display Messages

The high-performance D-Series Microscanner

 continuously monitors its 

ability to pro duce accurate temperature readings. If either the target’s  
temperature or the unit’s ambient temperature exceeds the operational 
limits, the beeper will sound one beep per second and the LED display will 
default to a display message.

For example, the D501 is designed to display temperature only if that  
temperature is within its specified measure ment range. If the target  
temperature is greater than 565°F (296°C), the display will read “HI.”

The battery voltage is also monitored. A low battery is indicated by a double 
beep each second. Temperatures will continue to be displayed as long as 
accuracy can be assured. If the battery drops below 5.7 volts the battery is 
considered “dead” and the display defaults to [----].

If a problem is discovered during the initial self-diagnostic test, the instru-
ment will display an error message (E-01), (E-02), etc. If any (E-##) error 
message is displayed, the unit will not function and must be returned to the 
factory for repair. (See section on Warranty for return instruction).

The following chart summarizes the fault conditions, and the  
associated indications:

Condition

Display  

Sound

Range

High Target

[HI]

1/sec 

>565°F (296°C) 

Low Target

[LO]

1/sec

<-65°F (-54°C) 

High Ambient

[HI A]

1/sec

>122°F (50°C) 

Low Ambient

[LO A]

1/sec

<32°F (0°C)

Low Battery

2/sec

<6.2v 

Dead Battery

[----]

2/sec

<5.7v

Processing Error

[E-##]

Return Instrument to Exergen for Repair

Содержание Microscanner D501

Страница 1: ...Instruction Manual Microscanner D Series D501 RS D501 High Performance Models...

Страница 2: ...our high performance Microscanner D Series features Full range resolution to 0 1 F or C SCAN MAX and MIN modes of operation A 10 second display lock An audible beeper to signal functional or condition...

Страница 3: ...initiate a new set of readings PUSH the power pushbutton at any time Operation Mode SCAN In the SCAN mode the target s instantaneous temperature is continuously displayed and updates 10 times per seco...

Страница 4: ...vides a marker with each unit It is only necessary to mark an area slightly larger than your Microscanner s nosepiece In the MAX or MIN mode you may want to check several points over an area In this c...

Страница 5: ...pment If electrical inspection is of particular importance contact Exergen for more informa tion relative to the dedicated electrical inspection instruments If the surface is metallic be sure to mark...

Страница 6: ...nose piece Although this will not harm the unit the indicated temperature may be a few degrees too high If this happens allow a few minutes between readings for the nosepiece to cool down and stabili...

Страница 7: ...a scan is indicated by the audible beeper Once a hot spot or cold spot is identified the exact temperature of that spot can be measured by touching the surface as previously ex plained When using the...

Страница 8: ...s This assures that the water on the wet membrane is evaporating at its maximum rate 3 With the Microscanner in the MIN mode quickly measure the temperature of the WET membrane immediately after you s...

Страница 9: ...p each second Temperatures will continue to be displayed as long as accuracy can be assured If the battery drops below 5 7 volts the battery is considered dead and the display defaults to If a problem...

Страница 10: ...ar ent material The shiny metal will reflect the transparent materi al s heat back to the Microscanner Wet or icy surfaces may interfere with the Microscanner s readings Water and other liquids are go...

Страница 11: ...nance required is to keep the sensor window clean This window located at the base of the nosepiece s reflective cup is made of a special crystal which passes infrared heat Dirt greasy films or moistur...

Страница 12: ...sure If after applying all of the above solutions the instrument s accuracy is still in question contact Exergen for further directions Warranty Exergen warrants the Microscanner D Series to be free o...

Страница 13: ...mm Repeatability 02 of absolute temperature Display Resolution 0 1 F or 0 1 C Display Update 10 times second Response Time Constant 60 mSec Field of View 1 1 53 degrees Spectral Sensitivity 2 to 20 mi...

Страница 14: ...Sensor Window Model D501 Nosepiece with Reflective Cup AECS Mode Selector Switch On Off Power Pushbutton High impact ABS Case Digital LED Display Microscanner D Series Features Overview...

Страница 15: ...Remote Sensor Probe Model D501 RS Microscanner D Series Features Overview...

Страница 16: ...Exergen Corporation 400 Pleasant Street Watertown MA 02472 U S A 617 923 9900 www exergen com 818098 Rev 3 Microscanner D Series...

Отзывы: