** If the surface is metallic, be sure to mark it as described in the
Measurement Tips section in Chapter Three.

2.4.3 High Temperature Targets

When using the DX™ to measure high temperature targets, above
100 °F, quickly touch the target's surface -- one second is long
enough to get an accurate reading.

Follow the procedures described above under 2.4.1 General
Operation (see pg. 5).  If the DX makes contact with a hot surface for
an extended time (10 seconds or more) the target will conduct heat.
Although this will not harm the unit, the indicated temperature may be
a few degrees too high.  Allow a few minutes between readings for
the nosepiece to cool down and stabilize.

If the DX makes contact with a hot surface and an error message [HI]
shows on the display remove the instrument from the object immediately.

2.5 MAX Mode Temperature Measurements

Use this measurement mode when making contact with the target or
when the target is large.  The DX is a wide-angle thermal scanner
with a fixed emissivity of 0.9.  The target's temperature is continuous-
ly displayed and updated 10 times per second for as long as you hold
the MEASURE button down. 

The DX displays the average temperature of everything within the
sensor's field of view measurement area.  The size of the measuring
area depends on the distance of the device from the target.  The DX
has a 1:1 field of view, which means that this unit will measure a 1"
diameter target area when the device is 1" from the target.

2.5.1 General Operation

For continuous temperature measurement of a target or for tempera-
ture measurement at a distance, use the following procedure.

1.

Determine the appropriate measurement distance from your
desired target.  Best measurements are made when the AECS
cup is in contact with the target.

2.

Make sure the MODE switch is on MAX.

3.

Point the nosepiece at the target.

4.

Press and hold the MEASURE button.

5.

Record the temperature from the display.

6.

Release the MEASURE button.  The last reading will remain
locked on the display for 10 seconds after releasing the button.

5

Содержание DX SERIES

Страница 1: ...EXERGEN 400 Pleasant Street Watertown MA 02472 Phone 617 923 9900 Fax 617 923 9911 www exergen com e mail industrial exergen com DX SERIES Handheld Infrared Thermometers User s Guide...

Страница 2: ......

Страница 3: ...perature Measurements 5 2 5 1 General Operation 5 2 6 SCAN Mode Temperature Measurements 6 2 6 1 General Operation 6 Chapter Three Measurement Tips 6 3 1 Optimizing Target Area Measurements 6 3 2 Meas...

Страница 4: ...e quality instrument calibration programs Operate the DX in SCAN mode for thermal scanning measurements over a range of tar get temperatures Thermal Scanning is very useful in many applica tions such...

Страница 5: ...00 F 86 to 871 C Unit includes a Remote Sensor 9 V DC Alkaline Battery and DX Instruction Manual Chapter Two Operation 2 1 DX Controls LED Display MEASUREMENT button LED Display Sensor Window AECS Cup...

Страница 6: ...ding the push button in turn the unit on with the MEASURE MENT button The instrument now reads in C instead of F To return to the original setting simply repeat the process 2 3 To Change Mode of Opera...

Страница 7: ...the button If the target is electrical equipment Exergen manufactures the Model E Microscanner instruments for safe and convenient inspection of electrical equipment Contact Exergen for more informat...

Страница 8: ...get is large The DX is a wide angle thermal scanner with a fixed emissivity of 0 9 The target s temperature is continuous ly displayed and updated 10 times per second for as long as you hold the MEASU...

Страница 9: ...se the following procedure 1 Determine the appropriate measurement distance from your desired target 2 Make sure the MODE switch is on SCAN 3 Point the nosepiece at the target 4 Press and hold the MEA...

Страница 10: ...Measurements for NIST Traceablity If you look at the DX s nosepiece you will see that it is really a small reflective cup with the sensor s window at its base When the DX s nosepiece touches the surf...

Страница 11: ...higher than 32 F 0 C because the instrument actually measures the air condensing on the ice a reac tion that generates a small amount of heat As long as the ice is not melting the DX will accurately r...

Страница 12: ...eadings If either the target s tempera ture or the unit s ambient temperature exceeds the operational limits the display will default to an error message The battery voltage is also monitored A low ba...

Страница 13: ...ion Error Code Remedy Measurement Fault High Ambient Temperature HI A Ambient 122 F 50 C Move to alternate location Low Ambient Temperature LO A Ambient 32 F 0 C Move to alternate location Low Battery...

Страница 14: ...ent Was the instrument s nosepiece in contact with the surface Touch the nosepiece flush to the target s surface and repeat the measurement If after applying all of the above solutions the instrument...

Страница 15: ...even Warranty Exergen warrants the DX to be free of defects in material and workman ship for a period of one year from the date of purchase and will repair or replace the returned instrument at no cha...

Страница 16: ...EXERGEN C O R P O R A T I O N 400 Pleasant Street Watertown MA 02472 617 923 9900 FAX 617 923 9911 818526 Rev 3...

Отзывы: