RME1-TC User’s Manual - Thermocouple Acquisition Module
Exemys
Table 11 - NETCONFIG Command
Command
Description
NETCONFIG:IP_address,Netmask
Gateway
Changes network configuration parameters. Netmask and IP
address equal to 0.0.0.0 enables DHCP
D.2.4. FACTRESET Command
At any time,the RME1-TC administrator can reset the original manufacturer’s configuration. This
option can be executed through the command console with the FACTRESET command, this way
you will have the original configuration.
Table 12 - FACTRESET Command
Command
Description
FACTRESET
Resets the original manufacturer’s configuration.
D.2.5. Reset Command
At any time you can restart the RME1-TC.
To restart the device you can execute the RESET command from the command Console.
Table 13 - Reset Command
Command
Description
RESET
Resets RME1-PT.
D.2.6 AINPUT Command
By means of this command you will be able to configure this type of thermocouple for each one
of the inputs.
From the command console you can execute the command AINPUT followed by the input
number to be configured (0-7) showing the type of thermocouple (0= Thermocouple B, 1=
Thermocouple E, 2= Thermocouple J, 3= Thermocouple K, 4= Thermocouple N, 5=
Thermocouple R, 6= Thermocouple S, 7= Thermocouple T)
Table 14 - AINPUT command
Comando
Descripción
AINPUTn:type
It changes the
type
of thermocouple for input n.
D.2.7 ALIST Command
From the command console, you can see all the analog inputs value by executing ALIST
command.
www.exemys.com Rev.
1
Page
37