8
EXCELL PRECISION CO., LTD.
01008044
ZSME100000117
3-4 G Value Adjustment
Steps:
1. Press 0 key and do not release, and then power on to enter int G value
adjustion mode.
2. Press T key to search the historic G value data.
3. Press Z key to set and confirm G value. (
→
0
←
for 16 key)
4. Press 0 ~ 9 key to input the value.
5. Press CE key to shift the flashing digit one space to left. (
1
C
1
for 16key)
6. Press
·
key to exit.
(1) The G value can be revise in 9 times.
(2) The G value range is from 9.78031 to 9.83217.
3-5 16key Function
3-5-1 Preset Using Conditions:
1. Preset function turn on (Note A)
2. The scale should be empty
3. There should not be general tare value (Note B)
(Note A)
Preset function turn on / off (SWA1 should be adjusted to ADJ)
Hold keypad C,it will shows Preset tare state after 2 seconds.
If Preset tare state is on, the Preset tare function will be off and it will restart-up.
If it is on the state of on, it show PT YES; If off, it shows PT NO
(Note B)
If Preset function is on, general tare function can be used (But the premise
should be CLEAR preset tare value), Preset tare function and general tare
function can not be used meanwhile.If you want to use preset tare function,
you should clear general tare value to use preset tare function.