Excelitas Technologies X-Cite Exacte Скачать руководство пользователя страница 2

 

 

 

X-Cite

®

 

exacte

: Quick Start Guide

 

 

 

 

SIDE 2 

 

Warning 

– Refer to Safety Precautions Booklet 035-00636R for all safety precautions prior to use. 

User Guide 035-00283 download: 

http://www.excelitas.com/Pages/Product/X-Cite-exacte.aspx

 

 

                                                                                              

 

Excelitas Canada Inc. 

2260 Argentia Road   Mississauga ON  L5N 6H7  

Tel: 1-905-821-2600 

 

 

Toll Free: 1-800-668-8752 

www.excelitas.com/x-cite 

 

Installing the Lamp Module 

1.

 

Be sure the AC power cord is disconnected from the unit. 

2.

 

Remove the screw from the lamp access side panel using the 3mm hex tool 

located on the bottom of the unit and remove the panel from the unit 

cover. 

3.

 

Carefully remove the lamp module from its container, holding only the 

ceramic component and glass rim 

4.

 

As illustrated below, position the lamp facing towards the front of the unit 

with the power leads facing towards you.  

 

Caution: 

Do not touch the bulb’s glass envelope or the inside surface of 

the reflector. Skin oils can cause the lamp module to fail prematurely.

 

 

5.

 

Make sure the middle of the lamp is in position to fit into the spring clamp. 

Slide the lamp until it snaps into the spring clamp. The leading edge of the 

reflector should fit snugly into the lamp holder recess. 

 

6.

 

Locate the 4-pin Intelli-Lamp sensor connector at the rear of the lamp 

module and connect it to its mate located on the top of the lamp-housing 

wall. 

Tip:

 the Intelli-Lamp connector will only attach in the correct orientation. If 

you are having difficulty attaching the connector, try rotating it by 180º. 

 

7.

 

Locate the Power connector with two leads and connect it to its mate 

located on the bottom of the lamp-housing wall. 

        

Tip:

 the Lamp Power connector will only attach in the correct orientation. If  

        you are having difficulty inserting the connector, try rotating it by 180º. 

 

8.

 

It is very important to ensure that the lamp anode cooling fin (bar) 

and lamp power lead, at the front of the lamp are centered within 

the lamp holder cutout.  Rotate the lamp as required. 

9.

 

Replace the lamp housing side panel and tighten the fastening screw. 

10.

 

Replace the hex tool in the clips underneath the system for future use. 

Note:

 if the lamp housing panel is not secured completely, the lamp will 

not strike and the “

bulb

” message will display when power is turned on to 

the unit. 

 

REGULATORY

 

(refer to user guide for complete details)

 

 

Product Safety and Electromagnetic Compatibility 

 

The X-Cite 

exacte

 has been tested and found to comply with product safety and 

electromagnetic compatibility requirements.  For a complete list of tests and for 

certification details, please contact your X-Cite representative or visit 

http://www.excelitas.com/Pages/Product/X-Cite-exacte.aspx

.   

 

CE Marking 

 

Council Directive 2014/35/EU 

  Low Voltage Directive 

 

Council Directive 2014/30/EU   

  EMC Directive 

 

Council Directive 2012/19/EU 

  WEEE Directive 

 

Council Directive 2011/65/EU 

  RoHS 

 

        

                              

                     

                   

 

WEEE Directive       

   China RoHS                  RoHS   

       FCC 

 

Rim 

*Center 
lamp power 
lead as 
shown