MFC
Maintenance
7.
211
7.5
Maintaining pH electrode
Clean and calibrate if there are fluctuations in the measured
values. A routine schedule cannot be given here because the
cleaning schedule depends heavily on the general condition of the
sample water. In general, calibrate approx. every 4 weeks.
Remove dirt on the glass membrane and diaphragm to prevent
measuring errors.
In particular, the ceramic electrode diaphragm may also be soiled
or coated with lime deposits.
Remove contaminants deposited on the surface of the membrane
glass; use diluted hydrochloric acid (up to 10 %), if necessary.
The electrodes should not be cleaned in a dry state because this
is more likely to smear the layer of dirt over the surface rather than
removing it. Under no circumstances may the membrane be
treated with abrasive cleaning agents.
The electrode must be rinsed subsequently with nothing other than
water.
Remove lime deposits on the glass membrane and the diaphragm
by immersing the electrode into hydrochloric acid (up to 10 %).
Rinse thoroughly with water or distilled water here as well; pH
electrodes age. This is often the cause for a slow display of the pH
value or a drop in the slope. pH electrodes typically last 1 to 2
years. However, routine maintenance of the electrodes
recommended.
7.6
Maintaining fluoride electrode
Routinely check the electrolyte level in the electrode (at least once
per week). The fill level should always be just under the filling hole,
approx. 25 mm above the sample water. Replenish the electrolyte,
if necessary.
Routinely calibrate the measuring system to guarantee safe
operation and accuracy.
Do not touch the glass surfaces with the cloth.
Содержание WALLACE & TIERNAN MFC
Страница 1: ...WALLACE TIERNAN MULTI FUNCTION CONTROLLER MFC from version V 2 00 INSTRUCTION MANUAL...
Страница 2: ...2 WT 050 580 000 DE IM 0714 MFC Please note Original manual...
Страница 6: ...6 WT 050 580 000 DE IM 0714 Introduction MFC 1...
Страница 40: ...40 WT 050 580 000 DE IM 0714 Description MFC 3...
Страница 225: ...MFC Complete Devices Retrofit Kits and Spare Parts 8 225 DEPOLOX 5 drawing non pressurised version W3T170031 W3T170031...
Страница 229: ...MFC Complete Devices Retrofit Kits and Spare Parts 8 229 DEPOLOX 5 drawing pressurised version W3T170032 W3T170032...
Страница 233: ...MFC Complete Devices Retrofit Kits and Spare Parts 8 233 VariaSens drawing non pressurised version W3T158742 W3T158742...
Страница 237: ...MFC Complete Devices Retrofit Kits and Spare Parts 8 237 VariaSens drawing pressurised version W3T158756 W3T158756...
Страница 247: ...MFC Certificates 10 247 10 Certificates...
Страница 248: ...248 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Страница 249: ...MFC Certificates 10 249...
Страница 250: ...250 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Страница 262: ...262 WT 050 580 000 DE IM 0714 Settings table MFC 11...
Страница 266: ...266 WT 050 580 000 DE IM 0714 Index MFC 12...