TO OPERATE WITH A BATTERY POWERED CONTROLLER

1. Turn the Battery Powered Controller 

upside down. Slide the battery cover 
in the direction of the arrow to 
remove the lid. Put in three (3) "AA" 
disposable alkaline or lithium batteries,
matching them to the plus (+) and 
minus (-) markings inside the Battery 
Powered Controller. Replace the cover 
on the Battery Powered Controller and 
slide the cover until it locks in place.
You can expect at least three (3) hours 
of operation from a new set of 
disposable batteries, (up to five (5) 
hours from high performance batteries).

2. Insert the nebulizer cable connector 

into the back of the Nebulizer Unit.
Slide the cable straight up the 
nebulizer body.

3. Connect the nebulizer cable to the 

Aeroneb

®

Go Battery Powered 

Controller.

4. Turn on the unit by pressing the "on/off" 

button on the Battery Powered 
Controller.The green light on the Battery
Powered Controller will light up. Visually 
verify that aerosol is coming out of the 
mouthpiece or mask to indicate that the 
device is functioning properly.

NOTE: (1)

If you are using batteries and the green light is flash-

ing, your batteries are low. You may finish your treatment, and 

7

12

DAILY: (Cont.)

1. Wash the Nebulizer Unit (Medication Cup and Cap, Nebulizer 

Body, and Mouthpiece or Mask Adapter and Mask) by hand 
in a solution of warm water and mild liquid dishwashing 
soap. Use soap intended only for washing dishes by hand.
Do not use dishwasher detergent as it is too abrasive. Do not 
leave parts soaking in soapy water for more than 20 minutes.

2. Rinse all pieces thoroughly under running hot tap water.

Make sure that no soap is left on any of the parts.

3. Shake off excess water and allow parts to air dry on a clean,

dry towel, out of the reach of children.

4. Make certain that all cleaned parts are completely dry before 

you store or use them. Do NOT remove the retaining ring 
under the medication cup.

NOTE:

If the Medication Cup is not properly cleaned after every

use the nebulizer may become clogged with medication.
Improper cleaning will result in poor performance and increased
time to nebulize the medication.

NOTE:

Water remaining on the underside of the Aeroneb

®

Go

generator after cleaning may cause the unit not to work. Be sure
the unit is completely dry prior to use.

NOTE:

Do NOT clean the nebulizer in a dishwasher.

OPTIONAL DISINFECTION:

If you would like to disinfect your Aeroneb

®

Go Nebulizer, you

may use the following method.

VINEGAR SOLUTION:

I. First, clean the Nebulizer Unit as described in the daily 

cleaning section, following steps 1- 2.

II. Fill a container with a mixture of three (3) cups hot water to 

one (1) cup white distilled vinegar.

(example: Heinz Distilled White Vinegar)

Содержание Aeroneb Go

Страница 1: ...ed or tampered with during the warranty period Shipping charges if any shall be paid by the purchaser NOTE There is no other express warranty Implied warranties including those of merchantability and fitness for a particular purpose are limited to the duration of the express limited warranty and to the extent permitted by law and all implied warranties are excluded This warranty does not cover pro...

Страница 2: ...tion Manual Thoroughly read and understand the information contained in these instructions before using the Aeroneb Go For future reference SAVE AND REFER TO THESE INSTRUCTIONS REPLACEMENT PARTS AND ACCESSORIES PART REORDER NUMBER 7000 Battery Controller with Cable 21105 7030 Wall Mount Adapter 21099 7070 Battery Controller with Cable 21105 Wall Mount Adapter 21099 7090 Battery Control Module 2221...

Страница 3: ...is manual Never operate this product if any of the parts are not working properly or have been damaged Do not use the Aeroneb Go Nebulizer in the presence of a flammable anesthetic mixture with air oxygen or nitrous oxide Do not use the Aeroneb Go Nebulizer with anesthetic breathing systems or lung ventilator breathing systems 17 2 PERFORMANCE Aerosol flow rate Greater than 0 25 mL per minute Resi...

Страница 4: ...eat a life threatening condition a backup device is necessary ELECTROMAGNETIC SUSCEPTIBILITY This device meets the requirements of IEC 60601 1 2 which describes requirements for electromagnetic compatibility EMC of medical devices This includes immunity to radio frequency electric fields and electrostatic discharge in addition to the other applicable requirements of the standard Compliance with EM...

Страница 5: ...f the low battery power light is flashing Replace the batteries or use the AC Powered Controller Check that the Battery Powered Controller cable is connected properly to the nebulizer and to the Control Module Connect the cable properly to the nebulizer and the Battery Powered Controller Refer to page 7 No visible aerosol when using AC Powered Controller Check that the AC Powered Controller is plu...

Страница 6: ...ct is being used to treat a life threatening condition a backup device is necessary In event of loss of electrical power remove the AC Powered Controller cable from the Aeroneb Go Nebulizer Unit and operate with the Battery Powered Controller refer to page 7 Never clean your Nebulizer Unit with solvents chemicals alcohol or abrasive cleaners Do not microwave any parts Do not use brushes sharp tool...

Страница 7: ...hirty 30 minutes IV Rinse all pieces thoroughly under running hot tap water V Shake off excess water and allow parts to air dry on a clean dry towel out of the reach of children VI Make certain that all disinfected parts are completely dry before you store them CONTROL lll GERMICIDE If you would like to disinfect your nebulizer you may use the medical disinfectant described below available from yo...

Страница 8: ... and Mouthpiece or Mask Adapter and Mask by hand in a solution of warm water and mild liquid dishwashing soap Use soap intended only for washing dishes by hand Do not use dishwasher detergent as it is too abrasive Do not leave parts soaking in soapy water for more than 20 minutes 2 Rinse all pieces thoroughly under running hot tap water Make sure that no soap is left on any of the parts 3 Shake of...

Страница 9: ...end of the cable after each use Clean the Nebulizer Unit and associated parts daily NOTE Removal and improper reinsertion of the retaining ring from the underside of the medication cup may cause the unit to leak or malfunction 11 8 CLEANING AND STORING YOUR AERONEB GO TO OPTIMIZE THE PERFORMANCE and life expectancy of your Aeroneb Go device we recommend you follow these cleaning instructions caref...

Страница 10: ...O OPERATE WITH AN AC DC POWERED CONTROLLER The nebulizer can be powered directly with an AC Powered Controller NOTE Aerosol mask is not provided 1 Insert the AC Powered Controller cable connector into the back of the Nebulizer Unit by sliding the cable straight up the nebulizer body 2 Plug the AC Powered Controller into a standard AC electrical outlet using the appropriate outlet adapter The green...

Отзывы: