evo Medical Solutions Aeroneb GO 7010 Скачать руководство пользователя страница 2

TO OPERATE WITH A BATTERY POWERED CONTROLLER

Model #7010

1.

Turn the Battery Powered Controller upside 
down. Slide the battery cover in the direction 
of the arrow to remove the lid. Put in three
(3) "AA" disposable alkaline or lithium batteries, 
matching them to the plus (+) and minus (-) 
markings inside the Battery Powered Controller.

Replace the cover on the Battery Powered Controller and slide the cover
until it locks in place.  You can expect at least three (3) hours of 
operation from a new set of disposable batteries, (up to five (5) hours
from high performance batteries).

2

. Insert the nebulizer cable connector into the 

back of the Nebulizer Unit. Slide the cable 
straight up the nebulizer body.

3

. Connect the nebulizer cable to the Aeroneb

®

Go Battery Powered Controller.

4.

Turn on the unit by pressing the "on/off" button 
on the Battery Powered Controller. The green 
light on the Battery Powered Controller will light 
up.  Visually verify that aerosol is coming out of 
the mouthpiece or mask to indicate that the 
device is functioning properly.

NOTE: (1)

If you are using batteries and the green light is flashing, your batteries

are low.  You may finish your treatment, and then replace the three “AA” batteries.

(2)

If the green light only comes on while depressing the button, immediately

replace the three “AA” batteries.

  

5.

Place the mouthpiece between your teeth, 
with your lips firmly sealed around the mouth
piece. If you use an aerosol mask, place the 
mask over your mouth and nose. Breathe in 
and out slowly through your mouth until 
aerosol formation stops. Do not cover ventilation
holes. 

NOTE: 

Aerosol mask is not provided.

NOTE: (1)

The Battery Powered Controller is designed to shut off after 16 minutes.  

If medication still remains in the cup, press the “on/off” button on the Battery
Powered Controller to continue your treatment.  

(2)

Battery life is dependent upon

not only the number of treatments per day, but also the length of each treatment.  
It is recommended that you always have a spare set of new 

batteries available to you.

6

. Press the "on/off" button on the Battery 

Powered Controller again to turn the nebulizer
off. Disconnect the cable from the Nebulizer 
Unit by sliding the cable straight down the 
handset, disassemble Nebulizer Unit and rinse 
Nebulizer Unit Parts after each use.  Do NOT

remove the retaining ring under the medication cup. Wipe dry the 
connector end of the cable after each use.  Clean the Nebulizer Unit 
and associated parts daily. Do NOT remove the retaining ring under the
medication cup. 

TO OPERATE WITH AN AC/DC POWERED CONTROLLER

Model #7030

1.

Insert the AC Powered Controller cable 
connector into the back of the Nebulizer Unit by 
sliding the cable straight up the nebulizer body.

2.

Plug the AC Powered Controller into a standard 
AC electrical outlet using the appropriate outlet 
adapter.  The green light on the AC Powered 
Controller will light up.  

3.

Visually verify that aerosol is coming out of the 
mouthpiece or mask to indicate that the device 
is functioning properly.

4.

Place the mouthpiece between your teeth, with 
your lips firmly sealed around the mouthpiece. 
If you use an aerosol mask,  place the mask over 
your mouth and nose.  Breathe in and out slowly
through your mouth until aerosol formation 
stops. Do not cover the ventilation holes during 
use.

5.

Unplug the AC Powered Controller from the wall
to turn it off. Disconnect the cable from the
Nebulizer Unit by sliding the cable straight 
down the handset, disassemble Nebulizer Unit 
and rinse Nebulizer unit Parts after each use. 
Wipe dry the Connector end of the cable after

each use. Clean the nebulizer unit and associated parts daily. Do NOT
remove the retaining ring under the medication cup. 

After completing Steps 1-4 on page 1, continue to follow the operating instructions that apply to your model.

TO OPERATE BATTERY POWERED CONTROLLER WITH AN
AC/DC ADAPTER

Model #7090

NOTE: 

This model has 2 modes of operation: 

Battery Powered Controller 

only

(see pg. 2)

OR

AC/DC with Battery Powered Controller.

1.

Plug the AC adapter into a standard AC 
electrical outlet, a green light on the adapter 
indicates that the AC adapter is turned on. 
Plug the AC cable into the Control Module, 
indicated by the Input Symbol (See 
Classifications and Markings chart in manual).

 NOTE:  

The AC/DC Adapter does not charge batteries.

2

. Turn on the unit by pressing the "on/off" button 

on the Battery Powered Controller.  The green 
light on the Battery Powered Controller will light 
up.  Visually verify that aerosol is coming out of 
the mouthpiece or mask to indicate that the 
device is functioning properly.

3.

Place the mouthpiece between your teeth, with 

your lips firmly sealed around the mouthpiece. 
If you use an aerosol mask,  place the mask over 
your mouth and nose.  Breathe in and out slowly
through your mouth until aerosol formation 
stops. Do not cover the ventilation holes during 
use.

4.

Press the "on/off" button on the Battery 
Powered Controller again to turn the nebulizer
off. 

5

. Disconnect the cable from the Nebulizer Unit 

by sliding the cable straight down the handset,
disassemble Nebulizer Unit and rinse Nebulizer
Unit Parts after each use.  Do NOT remove the 
retaining ring under the medication cup.  Wipe
dry the connector end of the cable after each

use.  Clean the Nebulizer Unit and associated parts daily.  

AC Adapter

Cable

To optimize the performance and life expectancy of 
your Aeroneb Go device,  follow the cleaning instructions
carefully as instructed on page 11 of your Aeroneb Go

Instruction Manual.

NOTE:

Removal and the improper insertion of the retaining ring

from the underside of the medication cup may cause the unit to leak
or malfunction.

NOTE: 

Please read all instructions before using.  Refer to your

Aeroneb

®

Go Nebulizer Instruction Manual for more complete

cleaning instructions.

IMPORTANT CLEANING

INFORMATION FOR ALL MODELS

To order replacement parts or accessories, visit your local home medical
equipment dealer or pharmacy

.

If you have any questions or problems regarding your Aeroneb

®

Go 

nebulizer, contact your local home medical equipment dealer or pharmacy. 

Your Aeroneb

®

Go was tested and inspected at the factory and is a premium

device that has a life expectancy of at least one year (730 treatments) with
proper care and cleaning. Lack of proper  care and cleaning, disassembly of
the retaining ring or misuse  will void the warranty.

IMPORTANT INFORMATION FOR ALL MODELS

2

3

4