background image

E2648-VOC

Rev 11.05.2021

The screwless quick connect spring terminals on the E2648 series devices are suitable for
a wide range of wires with a cross-section of 0.2...1.5 mm

2

. We recommend to strip the

wire end by 8...9 mm and use wire end sleeves.

To connect the wire, insert the wire end into the terminal hole. To disconnect, push the
spring-loaded terminal lever, pull the wire out, and release the lever.

Use a twisted-pair cable, e.g. LiYY TP 2×2×0.5 mm

2

or CAT 5, to connect the device to the

RS485 network. A and B on the device represent DATA+ and DATA- respectively, polarity
must be respected when connecting to an external RS485 network.

Both analog outputs can be independently changed between 4-20 mA and 0-10 V type
using jumpers J1 (OUT1) and J2 (OUT2). By closing pins on a specific jumper the related
output is 0-10 V, with an open jumper the output is 4-20 mA. Power restart is required after
changing the position of the jumpers.

NOTE! The outputs are not galvanically isolated from the external power supply and share
a common 0V. Allowed load resistance limits are stated in the Specifications table. To
power the instrument from an external power source, connect terminals 0V and +U to the
source. If the integrated mains power supply module is used, connect terminals L and N to
the mains.

NOTE! Actuator short-circuits should be avoided, to protect the instrument relays using
external fuses or safety switches.

18

Содержание PluraSens E2648-VOC

Страница 1: ...PluraSens Solvent Vapor Detector Transmitter E2648 VOC User Manual...

Страница 2: ...ing conditions 13 Installation guidelines 13 Mounting dimensions 14 Sensor probe handling 15 Electrical connections 16 Operation 19 Maintenance 19 Calibration 19 Delivery set 20 Order code for E2648 V...

Страница 3: ...E2648 VOC Rev 11 05 2021 Warranty 27 Manufacturer contacts 27 3...

Страница 4: ...ensity to air 2 0 Conversion 1 ppm 2 38 mg m3 Boiling point 56 11 C Low explosive limit LEL vol in air 2 5 Upper explosive limit UEL vol in air 12 8 Odor Characteristic pungent smell Hazards Highly fl...

Страница 5: ...Conversion 1 ppm 3 19 mg m3 Boiling point 80 C Low explosive limit LEL vol in air 1 2 Upper explosive limit UEL vol in air 7 8 Odor Hyacinth like odor Hazards Highly flammable Irritant Carcinogen May...

Страница 6: ...ow explosive limit LEL vol in air 3 3 3 Upper explosive limit UEL vol in air 19 Odor The characteristic smell of alcohol Hazards Highly flammable Gas air mixtures are explosive Inhalation of vapors le...

Страница 7: ...onversion 1 ppm 3 60 mg m3 Boiling point 77 1 C Low explosive limit LEL vol in air 2 Upper explosive limit UEL vol in air 11 5 Odor Sweet pear smell Hazards Flammable Short term exposure to high level...

Страница 8: ...Upper explosive limit UEL vol in air 6 75 7 1 Odor Characteristic chemical smell Hazards Highly flammable Gas air mixtures are explosive Inhalation possible effects irritation eyes nose lassitude wea...

Страница 9: ...s ortho xylene meta xylene para xylene Molar weight 106 Conversion 1 ppm 4 34 mg m3 Boiling point 144 4 C 139 C 138 35 C Low explosive limit LEL vol in air 0 9 1 1 Upper explosive limit UEL vol in air...

Страница 10: ...ply 11 30 VDC 24 VAC or 90 265 VAC as options Power consumption 2 VA Analog outputs 2 4 20 mA 0 10 V user settable Outputs assignment OUT1 2 gas OUT2 2 gas Outputs scale width 10 resolution Recommende...

Страница 11: ...sing contact 250 VAC 30 VDC 5 A max Alarm setpoints set release RE1 LOW 20 16 LEL RE2 HIGH 50 40 LEL RE1 LOW 100 80 ppm RE2 HIGH 300 240 ppm for 0 500 ppm range Other options Remote probe Protection I...

Страница 12: ...or 230 VAC as options for power supply module Remote probe The version of your detector is marked on the package If the symbol is marked on the equipment consult the documentation for further informa...

Страница 13: ...vier than air so the gas detector transmitter should be placed near the floor or potential leakage or formation source The device should be accessible for maintenance and repair The aforementioned con...

Страница 14: ...E2648 VOC Rev 11 05 2021 Mounting dimensions Front view Bottom view 14...

Страница 15: ...and water drops The filter should be replaced if it gets strongly contaminated To replace the PTFE filter unscrew the sensor head cap and remove the old filter Place a new filter into the cap and tigh...

Страница 16: ...ly and of the external devices Attach the power cable to the device without turning it on Using the connection diagram below connect the analog outputs and digital interface terminals to the relevant...

Страница 17: ...terminals OUT1 4 20 mA 0 10 V output OUT2 4 20 mA 0 10 V output 0V 0 V 24 VAC Neutral optional A RS485 A Data B RS485 B Data U 24 VDC 24 VAC Phase optional X5 terminals optional L 90 265 VAC Phase N...

Страница 18: ...to an external RS485 network Both analog outputs can be independently changed between 4 20 mA and 0 10 V type using jumpers J1 OUT1 and J2 OUT2 By closing pins on a specific jumper the related output...

Страница 19: ...Continuously on or off Make sure that the transmitter is properly mounted the external devices connected the power LED green on and the control LED red is constantly lit Make certain that the cable gl...

Страница 20: ...ng clamp for remote probe versions Order code for E2648 VOC options E2648 options Order code Remote probe 3 m cable E2648 VOC RP33 3 Remote probe 10 m cable E2648 VOC RP33 10 Integrated 90 265 V mains...

Страница 21: ...xible setting of analog output scales for each output Output zero and slope adjustment for calibration Free assignment of each of two relays to the chosen parameter Several relay control logic modes S...

Страница 22: ...ster Error codes 01 Illegal function 02 Illegal data address 03 Illegal data value 04 Slave device failure details of last error 04 can be read from register 0x0008 Modbus holding registers Register a...

Страница 23: ...as data 1 65535 512 0x00A7 167 40168 RW The change rate limit for gas units s 1 32000 0 no limit 0 0x00A8 168 40169 RW Integrating filter time constant s 1 32000 seconds 0 no filter 0 Address Reg MHR...

Страница 24: ...0 1000 s 0 0x00D9 217 40218 RW Minimal on off time for relay RE1 0 1000 s 0 0x00DA 218 40219 RW Minimal on off time for relay RE2 0 1000 s 0 Address Reg MHR RW Description Supported values dec Defaul...

Страница 25: ...0 1 in case the sensor is absent turn signaling off on analog output1 bit 3 0 1 in case the sensor is absent turn on signaling with low current high current on analog output1 if bit 2 0 this bit will...

Страница 26: ...efined 0x0107 263 40264 RW 0 value for analog output 2 32000 32000 gas units User defined 0x0108 264 40265 RW 100 value for analog output 2 32000 32000 gas units User defined The new value is applied...

Страница 27: ...Manufacturer will at its option either repair or replace a product that proves to be defective This warranty is void if the product has been operated in conditions outside ranges specified by the Manu...