E2618 series  transmitters  belong  to  the  PluraSens®  family  of  multifunctional
measurement  instruments.  The  instruments  utilise  gas  sensors  of  various  types  with
excellent repeatability, stability and long lifetime
E2618  series  provides  two  independent  analog  outputs  OUT1  and  OUT2,  user-
selectable to 4-20 mA or 0-10 V, proportional either to gas concentration or temperature.
RS485  Modbus  RTU  digital  communication  interface  allows  easy  instrument
configuration and integration into various automation systems.
A design with LCD indicator is available as an option.
The version of your transmitter is marked on the package.
If  symbol

  is  marked  on  the  equipment,  consult  the  documentation  for  further

information.
Safety requirements
Misuse will impair the protection of the equipment.

Always adhere to the safety provisions applicable in the country of use.
Do not perform any maintenance operation with the power on. Do not let water or
foreign objects inside the device.
External circuits connected to the equipment should have sufficient insulation rating
according to the environmental conditions and equipment power
A  disconnecting  device  that  is  marked  as  such  and  easily  accessible  should  be
included in the installation of this product.

Operating conditions
The device should be used in explosion-safe indoor areas, without aggressive gases in
the atmosphere. See Specification table for more details.
Installation and connections
Before  proceeding  with  the  installation  it  is  mandatory  to  read  carefully  the  Safety
requirements
 section and make sure to comply with all listed instructions. During the
installation of the detector the following points must be considered:

application (air quality control or leakage detection),
properties of the space under investigation (room geometry, direction and velocity of
air flows etc),
detected gas (relative  density to  air, temperature,  whether  the  gas is  flammable,  or
toxiс, or oxygen displacing),
avoid  strong vibrations, mechanical shock, and the sources of strong electromagnetic
interference,
the device should be accessible for maintenance and repair.

For  early  leakage  detection  install  the  sensor  as  close  as  possible  to  the  potential
leakage  sources  (flanges,  valves,  pressure  reducers,  pumps,  etc),  taking  into
consideration  other  points  listed  above.  For  general  area  monitoring  without  definite
leakage sources, the detectors should be distributed evenly in the room. For personal
safety control the detectors are installed in the breathing zone (at the height of the head
of people or animals). Recommended sensor position is vertical, pointing downwards.

See Installation guidelines section on the next page for more information.
1. Wall mount version: Unscrew four lid screws and detach the lid from the transmitter.
Attach the device to a wall with screws passing through mounting holes (for dimensions
see the antecedent drawing). Duct mount version: Cut hole with a diameter of 36...45
mm in the air duct at the chosen mounting place. Place the rubber flange aligning the
holes in the flange and the air-duct and fix the flange with four self-tapping screws. Pass
the sensor probe through the flange and adjust it to the appropriate depth. Unscrew four
lid screws and detach the lid from the transmitter.
2. Use M16 cable gland to let in cables of the power supply and of the external devices.
Attach  the  power  cable  to  the  device  without  turning  it  on.  Using  the  connection
diagram  below,  connect  the  analog  outputs  and  digital  interface  terminals  to  the
relevant devices according to your tasks.

The  screwless  quick  connect  spring  terminals  on  the  E2618  series  devices  are
suitable for a wide range of wires with cross-section 0,2...1,5 mm

2

. We recommend

to strip the wire end by 8...9 mm and to use the wire end sleeves.
To connect the wire, insert the wire end into terminal hole. To disconnect, push the
spring loaded terminal lever, pull the wire out, and release the lever.
Use twisted pair cable, e.g. LiYY TP 2×2×0,5 mm

2

 or CAT 5, to connect the device

to RS485 network. Use one pair for A and B wires and the second pair for common
0  V  and  power  +U  wires  to  connect  the  transmitter  to  Fieldbus  network.  Respect
polarity.  Overall  length  of  all  connections  via  RS485  interface  should  not  exceed
1200 m. The type of each analog output can be independently changed between 4-
20 mA and 0-10 V with jumpers J1 (OUT1) and J2 (OUT2).
With closed jumper the output is 0-10 V, with open jumper the output is 4-20 mA.
Power restart is required after changing the position of the jumpers.
The  output  assignments  and  scales  can  be  changed  by  Modbus  commands  (see
Annex 1 for more information).
By default both outputs OUT1 and OUT2 are assigned to gas concentration. The
device  has  built-in  temperature  sensor  which  may  be  tied  to  any  of  the  outputs.

NOTE  The  built-in  temperature  sensor  is  intended  for  temperature

compensation.  E2618  series  devices  are  not  meant  for  ambient  air  temperature
measurement.
We  recommend  to  set  the  difference  between  the  upper  and  bottom  limits  of  the
output scale not narrower than 20% of detection range (for CO detectors the scales
down to 5% of range are allowed). In any case, do not set the output scale below
the tenfold resolution of the device.

 NOTE If you use a version with LCD, only OUT1 is available.

3. Turn on the power. The sensor heating up takes ca. 60 seconds after switching on.
During this period analog outputs and Modbus interface are off. The operating status is
indicated  by  the  LED  on  the  PCB  of  the  device.  The  LED  response  to  different
processes is presented in the following table .

Process

LED mode

Sensor absence or malfunction

Blinking 0.5 Hz (90% off, 10% on)

Modbus response

The signal is modulated with short on-off pulses,
even single Modbus cycle is traceable*

Normal operating

Continuous light

4. Make sure that the transmitter is properly mounted, the external devices connected,
power on and control LED is constantly lit. Place the lid back and fix it with the screws.
The device is ready to use.
Sensor probe handling
The wall mount version of the transmitter is available with remote probe (see drawing
below for dimensions). The remote probe is connected to the main unit with shielded
cable. Default connection cable length is 3 m

The sensor probes of all types are equipped with a hydrophobic microporous PTFE filter
to protect the sensor from dust, dirt and water drops. The filter may be replaced if it gets
strongly contaminated. To replace the PTFE filter, unscrew the M25 nut and remove
the old filter. Place a new filter into the nut and tighten it again.

NOTE Never stab or press the filter near its centre where the sensor is located since

this may damage the sensor. 
The  recommended  orientation  of  sensor  probe  is  vertical  with  the  sensor  tip  pointing
downwards.  This  prevents  possible  accumulation  of  condensed  water  on  the  sensor
protection filter. The horizontal orientation is also suitable. Avoid upward position of the
sensor tip.
Configuring
Gas  transmitters  E2618  share  all  functionalities  of  the  PluraSens

®

  multifunctional

transmitter platform. The features and options include:
- digital output change rate limiting filter
- digital integrating (averaging) filter
- temperature measurement channel with internal sensor
- free assignment of each analog output to chosen parameter
- flexible setting of analog output scales for each output
- output shift and slope adjustment for calibration
A  standard  configuration  kit  includes  a  USB-RS485  converter  and  a  software  pack.
Please contact your Seller for more information.
Return to default settings
To reset the device’s Slave ID, baudrate and sbit number to factory settings, proceed as
follows:
1. De-energize the device

2. Connect the J3 jumper

3. Turn on the device

4. De-energize the device

5. Disconnect the J3 jumper

6. Turn on the device

RS485 communication interface
See Annex 1