Evikon PluraSens E2618-O2-L Скачать руководство пользователя страница 1

Oxygen

Colourless  gas,  one  of  the  main  components  of  the  air.  Oxigen  is  necessary  for
breathing  of  humans  and  animals,  oxygen  deficiency  is  harmful  for  health.
Strong oxidizer, sustains combustion.

Chemical formula

O

2

Molar weight

32

Relative gas density (to air)

1.1

Conversion

1 ppm =1.31 mg/m

3

Boiling point

-183°C

Odour

Odourless

Hazards

Strong oxidant. May reacts with
combustible and reducing materials
(oils, solvents etc), causing fire and
explosion hazard. Oxygen enriched
atmospheres (>22% O

2

) present a

significant fire and explosion risk.
Oxygen deficiency in air may lead to
loss of concentration, reduced
coordination, fatigue. At very reduced
levels fainting and death may occur.
Breathing of oxygen at increased
concentrations may lead to hyperoxia
(seizures, respiratory problems,
disorientation).

Normal atmosphere concentration

20.8...21%

Deficiency threshold (OSHA)

19.5%

Immediately life-treatening concentration <10%

Exposure limits

not established

Conversion of ppm to mg/m

3

 is calculated for 25°C and 1 atm.

Installation guidelines

(See 

Installation and connections

 section for general information.)

For  air  quality  control  install  the  detector  on  in  the  breathing  zone.  To  control  the
absence of oxygen (e.g. while working under inert atmosphere) place the sensor in the
working area.

Calibration

E2618-O2-L  transmitters  have  been  calibrated  with  standard  gas  mixtures  before
delivery.  Provided  that  the  sensor  is  used  under  moderate  conditions,  no  field
recalibration is required. Please contact your dealer for more information.

Maintenance

Do not perform any maintenance operation with the power on.
Clean the device with soft damp cloth. Do not use any abrasive cleaning agents. Do
not immerse the device into water or any cleaning media.

Delivery set

–Transmitter E2618 (wall mount or duct mount version)
– Mounting accessories:

4 screws with plastic dowels for wall mount version
rubber flange for duct mount version.
fixing clamp for remote probe version

E2618-O2-L_UM_EN 

         Rev 16.07.2019

Specifications

Sensor type

optical (based on fluorescence quenching)

Sampling method

diffusion

Typical detection range

0...25% vol

Resolution / digital unit

0.01 % vol

Response time T90

< 30 s

Sensor lifetime

>5 years

Calibration interval

maintenance free under moderate operating conditions

without extreme fluctuations

Signal update

every 1 second

Load resistance

R

L

 < (Us - 2 V) / 22 mA for 4-20 mA

R

L  

> 250 kOhm for 0-10 V mode

Digital interface

RS485, Modbus RTU protocol

no galvanic isolation

Power supply options

11...30 VD or 24 VAC

Power consumption

< 2 VA

Analog outputs

2 × 4-20 mA / 0-10 V, user settable

Outputs assignment

OUT1 2 gas; OUT2 2 gas

NOTE

 For LCD version only OUT1 is available.

Cable connections

screwless spring loaded terminals

Enclosure

grey ABS, wall or duct mount, protection class IP65

Dimensions

H82 × W80 × D55 mm

CE marking

according to 2014/30/EU and EN61326-1 requirements

Operating conditions

-30...60 °C, 15...90 %RH, atmospheric pressure ±10%

non ATEXrated areas, non-aggressive atmosphere

LCD option

Operating temperature

0...+50 °C

Display dimensions

72 × 36 mm

Number of digits

3.5 7-segment

Character height

14 mm

Other features

Backlight

Other options

Remote probe

Protection IP65, default cable length 3.0 m;

max height 80 mm, max diameter 65 mm

Warranty

This  product  is  warranted  to  be  free  from  defects  in  material  and  workmanship  for  a
period  of  one  year  from  the  date  of  original  sale.  During  this  warranty  period
Manufacturer  will,  at  its  option,  either  repair  or  replace  product  that  proves  to  be
defective. This warranty is void if the product has been operated in conditions outside
ranges  specified  by  Manufacturer  or  damaged  by  customer  error  or  negligence  or  if
there has been an unauthorised modification.

 

Evikon MCI OÜ

 

Teaduspargi 7, Tartu

[email protected]

 

Tel. +372 733 6310 

50411 Estonia

www.evikon.eu

PluraSens

®

Fluorescence-Based
Oxygen Transmitter

E2618-O2-L

User Manual