E2218
Rev 14.05.2021
9 - forced Modbus control,
value set in MHR (4)0205
0x00CB
(4)0204 RW
Forced value for analog
output 1***
0...1000 (0.0%...100.0% of
output scale)
0
0x00CC
(4)0205 RW
Forced value for analog
output 2***
0...1000 (0.0%...100.0% of
output scale)
0
0x00FF
(4)0256 RW
Sensor and analog outputs
status
bit [0] = 0 / 1 - sensor present /
absent, read-only!
bit [1] = 0 / 1 - analog outputs
deactivated / activated
bit[2]= 0/1 - if the sensor is
absent, turn signaling off/on at
analog output 1,
bit[3]=0/1 - if the sensor is
absent, turn on signaling with
low current/high current on
analog output1; if bit[2]=0, this
bit will be ignored,
bit[4]=0/1 -if the sensor is
absent, turn signaling off/on at
analog output 2
bit[5]=0/1 - in case of sensor
absent, turn on signaling with
low current/high current at
analog output2; if bit[4]=0, this
bit will be ignored,
bit[6]=0/1 - output 1:
current/voltage output,
read-only!
bit[7]=0/1 - output 2:
current/voltage output,
read-only!
bit[8]=0/1 - LED
deactivated/activated,
bit[9]=0/1 - buzzer
deactivated/activated,
bit[10]=0/1 - LED on/off
(normal measurement)
Notes: LED/buzzer
signalization (if activated):
in case of detectable sensor
absence or malfunction -
0.5Hz (90% off, 10% on) light
and/or sound signal;
in case of Modbus response
the light signal is modulated
with short on-off pulses, even
single Modbus cycle is visible;
15