background image

22

1

External View and Nomenclature

Finger hook

Stage plate

Specimen holder mount holes
(x 4)

Specimen holder

Tube

Seat
Same size as the stage plate.

Cup stage
Tilting limit angle : 30°

8-4  Cup Stage SZH-SC

The SZH-SC can be used only with reflected light illumination.

NOTE

2

Installation

Fit the cup stage seat into the stage plate mount hole of an applicable base.

Place the cup stage on the seat. Before placing, wipe the sliding surface on the cup stage and seat with a clean 
cloth.

Mount the stage plate.

Mount the specimen holder.

Clean the sliding surface periodically.

1

2

3

4

TIP

Before mounting, remove dirt and dust from the mount surfaces and handle carefully so as not to 
damage them.

STAD mount for ILLT SZX2-STADM is necessary for using the SZX2-ILLTQ/ILLTS LED transmitted light 
illumination base. For details, refer to the instruction manual for the SZX2-ILLTQ/ILLTS LED.

Stage plate

Stage plate mount hole

Specimen holder

Specimen holder mount hole

Cup stage

Applicable base

Seat

Sliding surface

1

3

4

2

NOTE

TIP

Содержание SZ51

Страница 1: ...ommend that you study this manual thoroughly before operating the microscope Retain this instruction manual in an easily accessible place near the work desk for future reference For details of product...

Страница 2: ......

Страница 3: ...3 Using the High Low Zoom Magnification Stopper 8 4 Using the Eye Shades 8 5 Mounting the Eyepiece Micrometer Disk 9 6 Using the Auxiliary Objectives 10 7 Selecting the Light Path SZ61TR 10 8 Adjusti...

Страница 4: ......

Страница 5: ...the black surface NOTE SAFETY PRECAUTIONS b a Hold the bottom of the stand with one hand while holding base column with the other hand to avoid tilting the microscope In case the specimen is damaged b...

Страница 6: ...that failure to follow the instructions in the warning could result in bodily harm to the user and or damage to equipment including objects in the vicinity of the equipment NOTE Indicates that failure...

Страница 7: ...age mounting holes M4 screw x 2 Accessory mount holes Eyepiece WHSZ10X WHSZ10X H WHSZ15X H WHSZ20X WHSZ20X H WHSZ30X H Body clamping knob Threaded for accessory mounting Microscope Body SZ61 SZ61 60 S...

Страница 8: ...mount Page 11 C mount clamping screw Confocality adjustment clamping screw Page 11 x 2 Light path selector lever Page 10 Pulled out Camera path 100 Right eyepiece sleeve light path 0 Pushed in Right...

Страница 9: ...on the stage plate Page 6 Adjust the interpupillary distance Page 6 1 2 Adjust the eyepiece diopter Page 7 This operation is not possible with the eyepieces without helicoids WHSZ10X 20X The operation...

Страница 10: ...y It is recommended to set the knob tension to a slightly higher level than the point where spontaneous drop occurs Hold the left and right focus adjustment knobs a with both hands fix the left knob a...

Страница 11: ...e into focus Place an easy to observe specimen on the stage plate Set the zoom adjustment knob c for low magnification look through the eyepiece with eyepiece micrometer disk and rotate the focus adju...

Страница 12: ...and the stopper ring on each zoom adjustment knob The right zoom adjustment knob is used to limit the higher end of the magnification and the left knob is used to limit the lower end Using an Allen w...

Страница 13: ...ring clockwise to attach it firmly Themicrometerdiskframemaybetightforcertainmicrometerdisks In this case turn the frame by holding the circumference with a light uniform force or by applying the fram...

Страница 14: ...Light Path SZ61TR a All of the auxiliary objectives are provided with the ESD protection performance The 0 3X and 0 4X auxiliary objectives cannot be used in combination with the standard base because...

Страница 15: ...lity e to focus on the monitor image After the adjustment tighten the clamping screws b and c Applicable cameras image pickup device size and field numbers The camera light path incorporates a 0 5X le...

Страница 16: ...focus adjustment knob does not rotate smoothly The rotation tension of the knob is set too high Decrease the rotation tension to an optimum level 6 6 The microscope body drops spontaneously causing th...

Страница 17: ...WHSZ30X H FN 7 24 mm eyepiece micrometer disk mountable Standard stand SZ2 ST Frame installation Mounting diameter 76 mm Focus adjustment Rack and pinion using ball guide Knob rotation tension adjust...

Страница 18: ...0 110AL1 5X 61 110AL0 5X 2 200 110AL2X 2 38 110AL0 62X 160 The working distances of the 110ALK0 3X and 0 4X can be varied according to the system The indicated magnifications 0 3X and 0 4X are the val...

Страница 19: ...touching glass surfaces NOTE To reserve the ESD protection performance be sure to ground the equipment using a grounding lead wire having a 4 mm banana plug The eyepieces are provided with theft preve...

Страница 20: ...cus adjustment knobs the microscope body can be installed in the opposite orientation to that shown in figure at the 180 rotated position TIP 3 Mounting the Stage Plate Mounting While applying the sta...

Страница 21: ...len wrench loosen the C mount clamping screw a and remove the C mount seat b Screw the C mount seat b into the camera c Place the C mount seat in the original position and tighten the clamping screw a...

Страница 22: ...o the SZ2 ET auxiliary sleeve when the SZ2 ST is used or the SZH P400 auxiliary column and the SZX R drop prevention ring also when the SZX2 ST or SZX2 illumination base is used 2 Applicable Bases and...

Страница 23: ...ng the Allen wrench to attach SZX STAD1 a to a base Attach U SRP or U SRG2 to SZX STAD1 a Attaching U SRP Place the positioning pin cylinder type c on the back side of the base and rotate the centerin...

Страница 24: ...Provided with SZH STAD1 Placing the notch a on the front side of the base secure the screws 2 pcs using the Allen wrench to attach SZX STAD1 b to a base Place the stage clamping knob c of BH2 SH hori...

Страница 25: ...ectly on the desktop STAD mount for ILLT SZX2 STADM is necessary for using the SZX2 ILLTQ ILLTS LED transmitted light illumination base For details refer to the instruction manual for the SZX2 ILLTQ I...

Страница 26: ...he seat Before placing wipe the sliding surface on the cup stage and seat with a clean cloth Mount the stage plate Mount the specimen holder Clean the sliding surface periodically 1 2 3 4 TIP Before m...

Страница 27: ...t the provided tube into the specimen holder to fix the container by pinching TIP Do not touch the sliding surface on the cup stage and seat with a hand If the sliding surface are contaminated wash wi...

Страница 28: ...EVIDENT EUROPE GmbH Caffamacherreihe 8 10 20355 Hamburg Germany EVIDENT EUROPE GmbH UK Branch Part 2nd Floor Part A Endeavour House Coopers End Road Stansted CM24 1AL U K OLYMPUS SCIENTIFIC SOLUTIONS...

Отзывы: