background image

Read Before Use

Model No.: EV-806

FOR THE

INSTRUCTION MANUAL

INSTRUCTION MANUAL

INSTRUCTION MANUAL

Comfy Stim

Digital Combination Unit

Distributed by:

Aug. 2019     V1.5

2460

Содержание Comfy Stim

Страница 1: ...Read Before Use Model No EV 806 FOR THE INSTRUCTION MANUAL INSTRUCTION MANUAL INSTRUCTION MANUAL Comfy Stim Digital Combination Unit Distributed by Aug 2019 V1 5 2460...

Страница 2: ...ead Wires 19 15 Lead Wire Maintenance 20 16 Electrode Options 20 17 Electrode Placement 20 18 Tips for Skin Care 21 19 Application of Reusable self adhesive electrodes 21 20 Adjusting the Controls 23...

Страница 3: ...oven way of treating muscular injuries It works by sending elec tronic pulses to the muscle needing treatment this causes the mus cle to exercise passively It is a product derived from the square wave...

Страница 4: ...APG protection Do not use it in the presence of explosive atmosphere and flammable mixture EMS 1 Federal law USA restricts this device to sale by or on the order of a physician ed by John Faraday in...

Страница 5: ...stimulators should not be used while driving operating machinery or during any activity in which involuntary muscle contractions may put the user at undue risk of injury Chapter 4 WARNINGS 1 The long...

Страница 6: ...TROL OUTPUT ON OFF SWITCH 3 PANEL COVER 4 LIQUID CRYSTAL DISPLAY 5 MODE CONTROL 6 SET CONTROL 7 INCREMENT CONTROL 8 DECREMENT CONTROL BACK BACK 9 BATTERY STRIP 10 BATTERY CASE 11 BELT CLIP SIDE 12 PRO...

Страница 7: ...dulation The pulse rate and width are automatically varied in a cycle pattern The pulse width is decreased by 50 from its original setting in 0 5 second then the pulse rate is decreased by 50 from its...

Страница 8: ...or Continuous Adjustable in 1 minute each step from 1 to 15 minutes and 5 minutes each step from 15 to 60 minutes Treatment time countdown automatically 22 Patient This unit can store 60 sets of oper...

Страница 9: ...nfor mation abou t this device model name serial number started with manufacturing year and week of the device supply voltage name of the manufacturer CE number and classification Please do not remove...

Страница 10: ...perceive individual pulses When using multiple electrode placement strategies such as com binations of point and contiguous electrode placements the quicker pulse rates are suggested Despite above re...

Страница 11: ...e also at the setting relaxation time Then the cycle starts over again Stimulation Contraction and Relaxation RAMP In order to achieve a comfortable exercise and avoid startle be cause of electrical s...

Страница 12: ...e treatment at home Chapter 18 TIPS FOR SKIN CARE To avoid skin irritation especially if you have sensitive skin follow these suggestions 1 Wash the area of skin where you will be placing the electrod...

Страница 13: ...use only 4 If irritation occurs discontinue use and consult your physician 5 Read the instructions for use of self adhesive electrodes before application Chapter 20 ADJUSTING THE CONTROLS 1 Panel Cov...

Страница 14: ...minutes and Continuous It can be adjusted by pressing the Set and Increment or Decrement controls The treatment time will countdown automatically in one minute increments Its output will be shut off w...

Страница 15: ...Set a EMS Program The settings can be adjusted according to the following steps f Turn on the Intensity After the electrodes are placed firmly on skin and the lead wires are plugged in the socket of d...

Страница 16: ...enu then press Increment or Decrement to adjust the setting If no instructions regarding the pulse width are given in therapy set the control to the suggested 70 120 us setting l Set Pulse Rate Pulse...

Страница 17: ...nual After being stored for 60 days or more the batteries may lose their charge After long periods of storage batteries should be charged prior to use BATTERY CHARGING 1 Plug the charger into any work...

Страница 18: ...s 2 Check the device for defective operating elements legibility of inscriptions and labels make sure the inscriptions and labels are not distorted 3 Check the usability of accessories patient cable u...

Страница 19: ...ial purposes The radio frequency emissions of the device are extremely low and in all probability do not cause any interference with other devices in the proximity It is recommended that you do not pl...

Страница 20: ...the electromagnetic environment specified below The customer or the user of the unit should assure that it is used in such an environment IMMUNITY test Conducted RF IEC 61000 4 6 Radiated RF IEC 61000...

Страница 21: ...the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electro magnetic propagation is affected by absorption and reflection from struc...

Страница 22: ...VERT 10 0V DIV HORIZ 2mS OUTPUT 57 1875Vpk pk Pulse Rate 100Hz Scope B VERT 10 0V DIV HORIZ 50mS Pulse Rate 5 000Hz Scope A VERT 10 0V DIV HORIZ 2mS OUTPUT 59 1825V pk pk Pulse Rate 148 8Hz Scope B V...

Страница 23: ...3 M MODE 50 Pulse Width Rate Modulation Load 500 ohm Pulse Rate 150Hz Pulse Width 300 s Scope C Modulation 50 3 SD1 MODE 40 Pulse Width Intensity Modulation Load 500 ohm Pulse Rate 150Hz Pulse Width 3...

Страница 24: ...dth 300 s Scope A VERT 10 0V DIV HORIZ 100 s OUTPUT 48 9975Vpk pk Pulse width 90 s Scope B VERT 10 0V DIV HORIZ 100 s OUTPUT 21 9957Vpk pk Pulse width 300 s EMS 1 S MODE Synchronous Load 500 ohm Pulse...

Страница 25: ...46 2 A Mode Alternate Load 500ohm Pulse Rate 150Hz Pulse Width 300 s Contraction Time 12 Sec Relation Time 12 Sec Ramp Time 6 Sec...

Отзывы: