PF9800 Series User's Manual

Copyright

©

EVERFINE, Copy or spread without authorization is prohibited.

8

Chapter 2 Specifications

2.1 Main specifications

(1) Input

Table 2.1

Input specifications

Item

Voltage

Current

All

PF9810

Rest

Range

600V/300V

/150V/75V

2000mA/800mA/

200mA/50mA

20A/8A

/2A/0.5A

Measuring range

3.0V~600V

0.5mA~2000mA

0.005A~20A

Input resistor

> 1M

<0.2

<0.02

Instantaneous maximum

allowable input for 1s

1000V

4000mA

40A

Continuous maximum

allowable input

700V

3000mA

30A

Range mode

Auto range

Range up

The reading value exceeds 110% of the rated range

Range down

The measuring value become

less than 30% of the rated range

The measuring value become less than

20% of the rated range

Frequency range

fundamental frequency is from 45Hz to 65Hz,brand width is 5kHz.

(2) Measurement accuracy

Table 2.2

Measurement accuracy of voltage, current and active power

Item

Conditions

Voltage

Current

Ac

tiv

e

po

we

r

Power

factor

Frequency

 

Содержание PF9800 SERIES

Страница 1: ...jiang Hi Tech Zone Hangzhou 310053 China Tel 86 571 86698333 Fax 86 571 86696433 E mail Global everfine net Globalservice everfine net http www everfine net Copyright EVERFINE Copy or spread without a...

Страница 2: ...ontents However should you have any question or find any errors please contact your dealer or EVERFINE sales office If you have different understanding to this manual please refer to R D department of...

Страница 3: ...lating the whole or part contents of this manual by any company or personnel without the written permission of EVERFINE is prohibited Otherwise it will be treated as infringement and the infringer wil...

Страница 4: ...ch the input interface when electrical circuit works 5 No pressure on the instrument and cables Keep the instrument far away from heat source 6 Please make sure that the ventilation on both upper and...

Страница 5: ...r and plastic bags to protect the instrument from temperature change shock and vibration during transportation 3 Please do not deposit the apparatus in the environment with dust smoke or chemical gas...

Страница 6: ...ear Panel 13 Chapter 4 Measurement Condition Settings 15 4 1 The Wiring Mode of the Measurement Circuit 15 4 2 Measurement Range 15 Chapter 5 Measurement Function 17 5 1 Display window 17 5 2 Function...

Страница 7: ...Failures and Resolution 28 7 1 Common failures and shooting 28 Chapter 8 Calibration and Inspection 29 8 1 Conditions for calibration and inspection 29 8 2 Technical documents of calibration and test...

Страница 8: ...er power factor voltage crest factor current crest factor harmonic analysis serial communication and other function The two digital power meters only have difference in the current measurement range P...

Страница 9: ...2 Instantaneous maximum allowable input for 1s 1000V 4000mA 40A Continuous maximum allowable input 700V 3000mA 30A Range mode Auto range Range up The reading value exceeds 110 of the rated range Range...

Страница 10: ...alue of voltage current and active power The relative values of total harmonic distortion of the voltage and current The calculation method of harmonic distortion is IEC the total calculation method o...

Страница 11: ...d power supplying input end should be insulated from each other with insulation resistance over 10M Withstand voltage AC2000V should last 1 minute between signal input end and the either of shell and...

Страница 12: ...panel of PF9810 PF9811 is shown in Figure 3 2 Figure 3 2 The Front Panel of PF9810 PF9811 3 1 1 Indicative Function and Operation Keys on the panel Table 3 1 Indicative function and Operation Keys on...

Страница 13: ...n cursor is added 1 when setting the value of upper and lower limit of power Table 3 2 Indicative function and Operation Keys on the panel of PF9810 PF9811 Functional classification Operation Keys Fun...

Страница 14: ...table of digital and character display 0 A K U 1 B L V 2 C M W 3 D N X 4 E O Y 5 F P Z 6 G Q c 7 H R h 8 I S i 9 J T u 3 2 Rear Panel The rear panel is shown in Figure 3 3 Figure 3 3 The Rear Panel o...

Страница 15: ...lack and A terminal is red 3 Communication interface PF9810 pf9811 Standard Configuration RS 232 Optional RS 485 LAN PF9804 This port is not a communication interface 4 Ground terminal To ensure safet...

Страница 16: ...ity must be used for these lines 4 2 Measurement Range 4 2 1 Range of voltage current and Power Table 4 1 The range of voltage current and power Range of voltage Range of current Range of Active Power...

Страница 17: ...n 20 of the current measuring range 2 The value of Upk or Ipk is less than 165 of the next measuring range 3 When the input effective voltage or current is more than 140 of the maximum range overrange...

Страница 18: ...F9804 Window Display Content Measurement Range A V U 3 0V 600V I 0 005A 20A P 1W 12 00kW PF 1 000 1 000 Hz 45Hz 65Hz B A C k W D PF Hz Table 5 2 The display content of PF9810 PF9811 Window Display Con...

Страница 19: ...PF9811 PF9810 When using the normal measurement the refresh rate of display is about 0 5 seconds and he refresh rate of display is about 2seconds by using the harmonic measurement 5 2 Function of Nor...

Страница 20: ...tor Power factor PF D Frequency Voltage Current frequency Hz Peak Ratio No for PF9804 Voltage Peak ratio CF A Current Peak ratio CF B Total Single harmonic measurement Total each harmonic voltage V A...

Страница 21: ...n are as follows 100 U U thd U n 1 k 2 k n 2 k 2 k or 100 I I thd I n 1 k 2 k n 2 k 2 k 100 U U U n 1 k 2 k k k or 100 I I I n 1 k 2 k k k Where Uthd or Ithd the total harmonic distortion of voltage o...

Страница 22: ...upper and lower limit the operations are as shown in Figure 5 3 Figure 5 3 The setup of PF9804 limit The Judgment of Upper and Lower Limit of Active Power and Current OK When the measured power and cu...

Страница 23: ...lt in alarm You can turn off the signal input and change the position of the connecting wires of the two terminals in rear panel Voltage and Current can not be changed at the same time which makes the...

Страница 24: ...PF9800 series User s Manual Copyright EVERFINE Copy or spread without authorization is prohibited 23 Figure 5 4 PF9810 PF9811 communication mode switching...

Страница 25: ...the installation is completed the file is placed on the target path C EVERFINE PF981X_X XX 6 3 Software Use The PC application software of PF981X is used to obtain the measured value and the data of h...

Страница 26: ...pyright EVERFINE Copy or spread without authorization is prohibited 25 4 Harmonic window display measurement data display area each harmonic measurement value as shown in Figure 6 2 5 Status Figure 6...

Страница 27: ...ave the current test data to the specified file 4 Print Output current display to the print 5 Print preview Can preview the paper after printing 6 Exit Sign out the software 6 3 2 Communication menu 1...

Страница 28: ...bservation Choose to display the main window or the harmonic window 6 3 4 Help menu 1 English Edition Set up the use of language for English After setting up you need to restart the software to take e...

Страница 29: ...is in the allowed range 2 Ensure that the display is not affected by noise interference 3 Check the test line connected correctly or not 4 Check the connection mode settings correct or not 5 Check th...

Страница 30: ...7 Digital multimeter calibrator 2 JJG 780 1992 AC digital power meter 3 JJG Zhe 89 2007 Electric parameters test meter 8 3 Equipments for calibration and test 1 Standard meter 0V 600V 0A 20A The preci...

Страница 31: ...ed such as wave crest ratio and power factor 8 4 2 Method of calibration and test 1 The wiring method of test is shown in figure 8 1 Please connect the instrument signal source and standard meter corr...

Страница 32: ...PF9800 series User s Manual Copyright EVERFINE Copy or spread without authorization is prohibited 31 Figure 8 1 Wiring diagram of calibration and test of PF9800 series...

Страница 33: ...9811 provide the same functionality in measuring voltage current power power factor frequency and harmonic analysis and serial communication the two models are just different in mesurement ranges Not...

Страница 34: ...0A Range Switching Mode Auto Switch Range Increasing The measured value is lower than 30 of therated range The measured value is lower than 20 of the rated range Range Decreasing The measured value is...

Страница 35: ...5 read Analysis number 1 50 Refresh rate Refresh every two second Communication PF9810 PF9811 Table 4 Communication Communication interface Standard configuration RS 232 or RS 485 CAN Baud rate 2400 C...

Отзывы: