31
Not all animals are being euthanized after the completion of gas
cycle:
1) Check that all connections in the entire system are properly attached.
2) Check that your gas supply has not run out.
3) Check that the flowmeter on the back of the control box is set to 50LPM.
4) It may be necessary to revise settings for the particular species that you are having difficulties
with. Contact Euthanex for information on this.
Personnel gas exposure limits are being exceeded:
1) Check that all connections in the entire system are properly attached.
2) Check that the blower box is properly wired to the controller.
3) Check that the exhaust/vent line is connected to the exhaust side of chamber.
4) Check that the 4” exhaust duct is properly connected between the exhaust blower and the in-
house exhaust system. The house exhaust must pull 100 – 150 CFM.
System not responsive:
1) Door message appears when doors are closed.
2) Screen Saver doesn’t disappear when screen is touched.
In both of the above please power off system at master power switch located on the lower left
of the rear side of the control box. Allow power to be off for at least 1 min.
If this troubleshooting guide does not resolve your problem, contact Euthanex Tech Support at