Operating Manual
alpha DO 1000
30
8
AUTO/MANUAL MODE
Regardless of the mode, you can control devices connected to Relay A or Relay B
from the front panel of this controller. In Automatic mode, the controller’s set point
values activate the relays. In Manual mode, you have manual control of the relays
so you can prime the pump or check pump status without operating the entire
system.
8.1
Auto Mode (Mode After Switch-On)
In this mode, the controller set-point values activate the relays. To view the set-
point values:
1.
Press the RELAY SELECTION (Rel A/Rel B) key
. The upper display
shows your set-point value for Relay A; the lower display shows “SP1”.
2. After two seconds the upper display shows your set-point value for Relay
B; the lower display shows “SP2”.
3. After an additional two seconds the controller will return to Measurement
mode.
8.2 Manual
Mode
In this mode, you can manually turn on and off the control devices connected to
Relay A or Relay B or both.
1. Press the RELAY CONTROL (auto/manu) key. The upper display shows
“000”; the lower display shows “S.Cd” to prompt you to enter the Advanced
Set-up code.
2. Press
the
▲
or
▼
keys to scroll the upper display until it reads “22”.
3. Press the ENTER key. The manual indicator by the RELAY CONTROL
key lights up.
Note: Pressing ENTER key at a value other than “22” will cause the
controller to revert to Measurement mode, and the relays will
remain in automatic mode.
4. Press the RELAY SELECTION key to select either Relay A or Relay B.
The LED next to the currently selected relay (A or B) will light.
5. The manual control options are now available.
6. If you selected Limit control: The upper display reads the current
measured value. The lower display shows “oFF” or “on” depending on the
relay status of the currently selected relay.
7. Press
the
▲
or
▼
keys to change the Relay on/off status. The LED
indicators at the right of the controller will also change between Red and
Green to indicate Relay status.