Instruction Manual
CyberScan pH1500
28
4.1.10 P10.0 Print Data
This setup option allows you to print current displayed data or data
stored in the meter’s memory to a computer or printer via its RS232
interface port. Note all the communication protocol for both the meter
and computer/printer must match before successful printing can be
performed.
To Select Print Data Option (Current / Memory)
1. Access the Select Print Data Option menu from any Measure
screen by pressing the
setup
key.
2. Use
the
▲
or
▼
key to scroll until the screen display as shown.
save
view
press
to set value
press
to accept
enter
Data
3. Press
enter
key to make selection and use the
▲
or
▼
key to
toggle between Current or Memory data print out selection. Press
enter
for the desired selection OR press
std
to exit from this
page.
4. Press
enter
key to confirm selection.