Instruction Manual
COND 6+, TDS 6+, SALT 6+
-12-
NOTES:
1.
To protect from erroneous calibrations, the allowable
tolerance is ±40% of the factory default value. If calibration is
attempted with standards that fall outside this tolerance
range, the error message “
Err 1
” is indicated and meter will
return to measurement mode. For example, a 40% tolerance
of a 1413 µS/cm standard, is 848 µS/cm to 1978 µS/cm.
2.
If the measured temperature (°C) of the calibration solution is
below 0 °C or above 50 °C, the error message “
Err 2
” is
indicated and meter will return to measurement mode.
3.
Low conductivity standard solutions (less than 20 µS/cm) are
unstable and are very temperature dependent. As a result,
reproducible calibration results are challenging in lowest
measurement range (0.00 to 20.0 µS/cm).