
15
8.2
Setup mode overview
P1.0: Set Point 1
P1.1: select relay 1 set point value
P1.2: select relay 1 as low or high set point
P1.3: set relay 1 hysteresis value
P2.0: Set Point 2
P2.1: select relay 2 set point value
P2.2: select relay 2 as low or high set point
P2.3: set relay 2 hysteresis value
P3.0: Range
P3.1: select measurement range (with the corresponding cell constant value).
Eight measurement ranges are available.
P4.0: Temperature Data
P4.1: select ATC or MTC
P4.2: select temperature coefficient value
P4.3: select normalization temperature
P5.0: TDS Calibration points
P5.1: view TDS value at which calibration was performed
P6.0 Electrode Properties
P6.1: view cell constant value after calibration
P6.2: view temperature offset value after calibration (only if in ATC mode)
P7.0: TDS Factor
P7.1: select TDS factor for appropriate application
P8.0: Reset
P8.1: select yes/no to reset controller to factory defaults