ADVANCE D SETUP FUNCTIONS
The advanced setup mode lets you customize your transmitter’s preferences and
defaults. This transmitter features different sub groups that organize all setup
parameters.
The sub-groups are:
6.1
Offset Setting (Unavailable in ORP mode)
As a result of the continuous measurement and the application concerned, it may
not be convenient to remove the electrode for calibration. In such cases, an on-line
Offset adjustment is offered.
1. Press ENT key. Transmitter displays ‘
OFS
’. Press
ENT again. Upper display shows amount of
offset, in terms of pH; lower display shows current
measured pH value.
SETUP
HOLD
SETUP
HOLD
2. Use
▲
or
▼
keys to introduce offset to pH
readings. As the lower measured reading
changes, upper display records the amount of
offset. Maximum of ± 2.00 pH offset is allowed.
pH
READY
.
3. Press
both
▲
or
▼
keys together to return to ‘
MEAS ‘ mode.
.
pH