Instruction Manual
α
lpha-pH2000D
- 28 -
6.8
Wash relay (WASH) sub-function
In this sub-function you set the parameters for the wash relay.
1
2
2
3
3
4
5
ENT
ENT
ENT
ENT
ENT
ENT
ENT
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
off
1.
Select the “WASH” sub-function, and then press the ENT key.
2.
Enabling/disabling wash function: press the
▲
or
▼
key to enable (WASH ON) or disable
(WASH OFF) wash function. Press the ENT key to confirm your selection.
3.
Selecting the relay status condition: press the
▲
or
▼
key to choose desired relay status
(de-energized = “DEEN” or energized = “EN”). Press the ENT key to confirm your selection.
4.
Setting the wash interval in hours: Press the
▲
or
▼
key to set the desired wash interval
(setting range: 0.1 to 200.0 hours). Press the ENT key to confirm your setting.
5.
Setting the wash duration in seconds: Press the
▲
or
▼
key to set the desired wash
duration (setting range: 1 to 2000 seconds). Press the ENT key to confirm your setting.
Note: During wash cycle the transmitter is set to HOLD. For safety reasons, the
HOLD function is activated 5 seconds prior and deactivated 10 seconds after the
wash cycle.
Continue with Setup mode procedures, or return to Measurement mode by pressing the
▲