68X256903  Rev  1  10/99

Before Calibration

Remove cover of Transmitter and ensure that electrode is correctly
connected as above. Connect Amp meter to the 4/20mA output of
Transmitter (Appendix 1).

Dissolved Oxygen Transmitter Calibration

1) 

Short Dissolved Oxygen input and observe Amp meter display;

2) 

Amp meter should read 4mA; if not, calibration is necessary;

3) 

Locate “Offset adjustment” trimmer; Using a fine screwdriver,
slowly turn trimmer till Amp meter reads 4mA

;

4) 

For greater accuracy, adjust the slope using Air as a reference;

5) 

Hold the dry probe in Air;

6) 

In 100% Saturation (in Air), output is 12 mA

;

7) 

In 200% Saturation, output is 20 mA

;

8) 

Using a fine screwdriver, slowly turn trimmer, marked ‘Slope’, till
Amp meter reads correct mA value with respective Oxygen
Saturation levels.

9) 

Remember to disconnect Amp meter and close cover.

NB: For other % Saturation values, expected current output is based on:

mA= {(16/200) x % Saturation of Dissolved Oxygen} + 4mA

Specifications

Model

ECDOXTR

Range

0 to 200%

Accuracy

±

 1.5% Full Scale

Calibration

1 point (100% air saturation)

Input

Only with Galvanic DO probe – 2 wires

Output

4-20 mA

Operating
Voltage

DC 12 V to 24 V

Load

100 Ohms max. for 12 V;

600 Ohms max. for 24V

Housing

77 x 77 x 28 mm, field mountable

Appendix 1

Connecting Amp Meter for Calibration

Figure 3 - Method One

Figure 4 - Method Two

10  Ordering Information

pH

Item

Cat Code

pH Transmitter

ECPHXTR

Combination pH Electrode, General, integral 5m cable

ECARTSO05B

Combination pH Electrode, w/PMP, integral 5m cable

ECARGTSO05B

Dissolved Oxygen

Item

Cat Code

Dissolved Oxygen Transmitter

ECDOXTR

General Purpose D.O. probe, 0.5 – 40 ppm, Industrial,
integral 5m cable

ECDOGEN

General Purpose D.O. probe, 0 – 10  ppm, Industrial,
integral 5m cable

ECDOTPll

ORP

Item

Cat Code

ORP Transmitter

ECORPXTR

Combination Platinum ORP Electrode, w/PMP,
integral 5m cable

ECHTPTTSO05B

Combination Gold ORP Electrode, w/PMP, integral
5m cable

ECHTAUTTSO05B

Conductivity

Item

Cat Code

Conductivity Transmitter

ECCONXTR

2-Pin SS,k=1.0, integral 3m cable

ECCONSEN46

Temperature

Item

Cat Code

Temperature Transmitter

ECTEMPXTR

Pt 100 probe, SS316

ECPT10001M or

ECPT10005M

Power Supply

Item

Cat Code

120V Power Supply

EC-120-XTR

220V Power Supply

EC-220-XTR