L30C-30A-UK-02-01 

May 2009 

Call service:

 Is displayed when the service timer has counted down to 1, if the timer 

is  active.  The  control  is  fully  operational  and  the  service  timer  can  be  reset  by  our 
service technicians. 
 

Device blocked:

 If displayed, contact EUROWATER for further information. 

 
 

Low  battery:

  Is  displayed  when  the  internal  battery  must  be  changed.  Contact 

EUROWATER for further information. 
 

Parameters:

 Enter code 111 or 222 to access plant parameters. The codes can be 

changed. 
 

Parameters:

  When  correct  code  has  been  entered  access  is  given  to  the  con-

figuration menu. The plant's program version is displayed in the second display line. 

 
 

6.2  

Parameters – RO

 

 
 
 
 
RO, limit value:

 Set the highest acceptable permeate conductivity of the RO plant. If 

the value set is exceeded, the plant carries out a quality rinse and then puts on an 
alarm. 
 

RO, pre-rinse:

 Set the duration of pre-rinse. Value is set as whole minutes and can 

be from 1 to 59. 
 

RO, q. rinse:

 Set the duration of quality rinse. Value is set as whole minutes and can 

be from 1 to 59. 
 

RO,  post-rinse:

  Set  the  duration  of  post-rinse.  Value  is  set  as  whole  minutes  and 

can be from 1 to 59. 
 

RO,  stand-by:

  Set  the  duration  of  stand-by  time.  Value  is  set  as  whole  hours  and 

can be from 1 to 45. Stand-by time is the interrupted time in which the plant is not in 
operation.  If  the  plant  is  out  of  operation  longer  than  the  stand-by  time,  a  rinse  is 
performed. 
 

RO,  rinse:

  Set  the  duration  of  rinse  after  stand-by  time.  Value  is  set  as  whole 

minutes and can be from 1 to 59. If during rinse the RO plant receives a start signal, 
the rinse is completed before the plant starts. 
 
 

 6.3 

Parameters - softening

 

 
 
 
 

Softening, backwash:

 Set duration of backwash. Value is set as whole minutes and 

can be from 1 to 59. 
 

Softening, draw:

 Set duration of brine draw. Value is set as whole minutes and can 

be from 1 to 59. 
 

Softening,  rinse:

  Set  duration  of  rinse.  Value  is  set  as  whole  minutes  and  can  be 

from 1 to 59. 
 

Softening, q. rinse:

 Set duration of quality rinse. Value is set as whole minutes and 

can be from 0 to 99. Quality rinse is only possible with two-tank plants. Just before a 
tank is commissioned, a rinse of the tank is performed to ensure good water quality. 
If value is set to "0", the rinse is not carried out. 
 

Softening,  capacity:

  Set  plant  capacity  per  tank.  Setting  depends  on  the  water 

meter chosen. If the water meter counts 10 or 100 litres/pulse, the value is changed 
to 0.1 m³. If the water meter counts 1000 litres/pulse, the value is changed to 1.0 m³. 

Содержание SE 30

Страница 1: ...24 VDC Allowed supply voltage 20 4 28 8 VDC Maximum current consumption 80 mA at 24 VDC Digital input 12 Digital output direct 9 Maximum current reserve 1 A Voltage 20 4 28 8 VDC Digital output potential free 5 Maximum current 1 A Maximum voltage 30 VDC Total maximum current outputs 5 A Analog input 2 4 20 mA Analog output 1 4 20 mA Protection class IP54 Ambient temperature 0 50 C Storage temperat...

Страница 2: ...C 30A UK 02 01 1 May 2009 2 ELECTRICAL CONNECTION AND INSTALLATION RUN CPU is active SC CPU error I0 I11 Digital input O0 O15 Digital output F1 fuse for CPU T1A 5 x 20 mm F2 fuse for outputs T5A 5 x 20 mm ...

Страница 3: ...valve T1 11 12 I5 Block automatic regeneration 35 36 O5 Outlet valve T2 13 14 I6 Pressure switch 37 38 O6 RO inlet valve 15 16 I7 Block operation 39 40 O7 RO quality rinse valve 17 18 I8 RO pump in operation 41 42 O8 RO pump 19 20 I9 RO protective motor switch disconnected 43 44 O9 Alarm potential free 21 22 I10 Conductivity OK 45 46 O10 Dosing potential free 23 24 I11 Alarm reset 47 48 O11 Output...

Страница 4: ...red value on the numeric keypad Press ENTER to confirm your entries Special functions can be enabled or disabled These functions are displayed with when enabled and when disabled Press ENTER to select a function and the symbols blink Now press ENTER again to confirm your selection If you want to disable a function the procedure is the same If the control is in configuration menu and the keypad has...

Страница 5: ...L30C 30A UK 02 01 4 May 2009 ...

Страница 6: ...n as text on the display and as red LED alarm Press ENTER to acknowledge an alarm Regeneration of the softening plant can be started manually by depressing key 0 and at the same time pressing down key 1 for approximately two seconds It is also possible to step through the phases by depressing key 0 and pressing down key 2 for approximately two seconds If the RO plant is in start up it is also poss...

Страница 7: ...as been so for so long that a planned stand by rinse is performed 10 minutes and 00 seconds of the rinse remain Softening operation The softening plant is in operation and can supply 9 9 m of softened water before regeneration Tank 1 is the next to be regenerated The character before T1 is an indicator of the pulse water meter When a pulse is registered the character changes to vertical for a shor...

Страница 8: ...59 RO stand by Set the duration of stand by time Value is set as whole hours and can be from 1 to 45 Stand by time is the interrupted time in which the plant is not in operation If the plant is out of operation longer than the stand by time a rinse is performed RO rinse Set the duration of rinse after stand by time Value is set as whole minutes and can be from 1 to 59 If during rinse the RO plant ...

Страница 9: ...nge for the analog outlet Only calibration of the analog output signal is possible 4 mA point Set the desired value applicable to 4 mA on output Possible setting from 9 9 to 25 µS cm 20 mA point Set the desired value applicable to 20 mA on output Possible setting from 25 to 75 µS cm Number of tanks in softening plant Set the number of tanks on the softening plant Possible setting 1 or 2 If one tan...

Страница 10: ... pressing the keys 0 and 1 or 0 and 2 respectively Softening reset capacity If regeneration of the softening plant is started manually the capacity is not reset automatically It is possible to do so manually Factory setting The set values are changed to factory setting Select language Among these languages select the desired display language 01 English UK 02 German DE 03 Danish DK 04 French FR 05 ...

Страница 11: ...L30C 30A UK 02 01 10 May 2009 ...

Страница 12: ...y of our own make We continuously supply spare parts for more than 25 year old plants Consumables Water softening salt Test kits for periodical hardness control after the softening plant Filter media for pressure filters Ion exchange resins Membranes for reverse osmosis plants Filter bags UV lamps Temporary Needs In case of emergency or temporary needs we offer a wide range of rental units All uni...

Отзывы: