PROBLEM 

POSSIBLE SOLUTION 

POWER led does not turn on (the compressor 

is not working) 

Check the cable, the plug and the socket; 

check compressor fuse. 

The mattress does not inflate or only partially 

inflates 

Check tubes, connections, the lateral 

distribution band; if the problem continues to 

exist, please contact maintenance service. 

ALARM led remains turned on 

Check tubes, connections, the lateral 

distribution band; if the problem continues to 

exist, please contact maintenance service. 

When setting the compressor to the minimum, 

the alarm activates 

This setting is too low for patient’s weight. 

Increase comfort level. 

Patient sinks into the mattress without alarm 

going off or the mattress is too rigid 

Adjust comfort level as explained in the 

previous paragraph. 

 

 

 

 

 

2

2

0

0

.

.

 

 

 

 

 

 

 

 

T

T

E

E

C

C

H

H

N

N

I

I

C

C

A

A

L

L

 

 

C

C

H

H

A

A

R

R

A

A

C

C

T

T

E

E

R

R

I

I

S

S

T

T

I

I

C

C

S

S

 

 

 

System 

Euro Basic 

Type of mattress 

Overlay mattress 

Dimensions – overlay mattress (inflated) 

200 x 85 x 10,5 cm 

200 x 85 x 15 cm (with the addiction of PU foam base) 

Inflatable elements 

Nylon-PU -  depth: 10 cm 

Circular section 

Maximum patient weight (Kg) 

150 

Emergency deflation time (CPR) 

10 seconds 

Inintial inflation time 

25 minutes 

Cover  

Waterproof PES/PU , transpirant to water steam 

Compressor 

shock proof  ABS 

enclosure 

30 x 18 x 13,4 cm 

Содержание EURO BASIC

Страница 1: ...User Guide Euro Basic Rev 06 del 11 2011 pag 1 di 16 User Guide Antidecubitus systems EURO BASIC...

Страница 2: ...nded use 5 6 Improper and incorrect use 5 7 Warnings 6 8 Control panel 7 9 Installation 7 10 Starting the compressor 8 11 Dynamic and static functioning 8 12 Nursing 8 13 Inflation regulation 8 14 Ala...

Страница 3: ...hanged or updated This booklet is given with the product It is part of it and it is up to the customer to preserve it properly 2 2 S ST TA AN ND DA AR RD DS S A AN ND D D DI IR RE EC CT TI IV VE ES S...

Страница 4: ...ot dispose of as normal waste Back of the compressor Characteristics of the fuse Year of production S N Serial Number Base of the mattress and bottom of the cover Manuifacturer 4 4 D DE ES SC CR RI IP...

Страница 5: ...RO OP PE ER R A AN ND D I IN NC CO OR RR RE EC CT T U US SE E Improper and incorrect use description Consequences remarks Using the system in order to prevent decubitus ulcers on patients who run a hi...

Страница 6: ...ctionality 7 7 W WA AR RN NI IN NG GS S Keep system away from flames heaters and protect from humidity Avoid any contact with flammable gasses and liquids RISK OF EXPLOSION Before managing the compres...

Страница 7: ...as to guarantee patient s safety conditions Make sure that CPR valve near the head end is shut by pulling the bigger red block firmly without pushing the transparent flap and making sure not to extrac...

Страница 8: ...ive elements alternatively apart from the three head elements which always remain static The dynamic cycle is programmed to deflate one of three sections every three minutes plus 1 minute when all thr...

Страница 9: ...n Check that connection hoses are not kinked twisted or entangled to the bed check that inflatable elements are clicked into the clutches of the lateral distribution band If the alarm continues please...

Страница 10: ...essary add delicate alcohol free bleach free or chlorine free disinfectant Machine wash max temperature 60 C using delicate alcohol free bleach free or chlorine free disinfectant tumble drying max tem...

Страница 11: ...part of the compressor With the aid of a screwdriver extract blown fuse and replace it with a new one with the same electrical characteristics Use only 5 mm 250V delayed 500 mA fuses 1 18 8 H HA AN ND...

Страница 12: ...ution band if the problem continues to exist please contact maintenance service When setting the compressor to the minimum the alarm activates This setting is too low for patient s weight Increase com...

Страница 13: ...f may need in order to repair the compressor IMPORTANT for qualified staff before handling compressor disconnect the power cord All maintenance must be undertaken by qualified staff only Please contac...

Страница 14: ...ith 3 The manufacturer will decide whether to repair the fault or to substitute the faulty part with a new one 4 The warranty does not cover faults deriving from incorrect installation or use faults d...

Страница 15: ...User Guide Euro Basic Rev 06 del 11 2011 pag 15 di 16...

Страница 16: ...User Guide Euro Basic Rev 06 del 11 2011 pag 16 di 16 Made by rev 06 11 2011...

Отзывы: