Clear light™ Makes Professional IPL Treatment Easier and more Efficient
22
(11) Current shot count
Display the number of current shot.
4.3.5 State display
(11) (12) (13)
(14) (15)
(12) Current shot erase button
A button for erasing current shot.
(13) Total shot count
Display the number of total shots.
Press the button to show the number of total shots for 2 second.
(14) Temperature display
Display current temperature of filter.
(15) Cooling button
Button for cooling operating. It is automatically operating when power on and ready
mode and the button will be turn to red.
* Cooling automatic system
<Control 1> -5℃ ~ 0℃, <Control 2> 0℃ ~ 15℃
- When the power is on, it is automatically start cooling on and operate <Control 1>
system to cool down to -5℃.
- If it doesn’t have any input for 1 minute, <Control 2> system will be start.
- When it have an operating commend, the cooling will be operate automatically.
ON
OFF