![ETT Olfactometer 2 Скачать руководство пользователя страница 22](http://html.mh-extra.com/html/ett/olfactometer-2/olfactometer-2_manual_2434540022.webp)
22
© Emerging Tech Trans, LLC
Figure 29 - Distribution of odorant bottles in the odorant carrier.
Connecting the tubing
With your device, you will also receive a bound bundle of Teflon tubing (length 8 m) and a single line
(length 2 m) that will connect the subject applicator mask to the odorant carrier output. To prevent
damage and/or premature wear to both the quick release connectors as well as the Teflon tubing always
make sure to push the tubing in straight without bending it until the tube is all the way into the
connector. The tubing is released by first pushing the blue or black ring sleeve on the connector into the
connector as far as possible. While holding the sleeve into the connector one can now safely pull the
Teflon tubing out of it without damage (Figure 30).
Figure 30 - Connect and release the provided Teflon tubing without damage to both the connector and the tubing
Make sure to align tubing between the device and the odorant carriers as labeled on the tubing via color
code (older devices) or numerical code (recent versions) as shown in Figure 31 and Figure 33.
Содержание Olfactometer 2
Страница 1: ...ETT Olfactometer ETTDirectControl The First Paradigm Manual and Tutorial Revision 2021 1 ...
Страница 2: ...2 Emerging Tech Trans LLC ...
Страница 52: ...52 Emerging Tech Trans LLC Odorant flow setup Carrier Setup Check valve layout in the odorant carrier ...
Страница 53: ...53 Emerging Tech Trans LLC Typical air flow configuations ...
Страница 54: ...54 Emerging Tech Trans LLC ...
Страница 55: ...55 Emerging Tech Trans LLC Configuration for multiple channels at once ...
Страница 56: ...56 Emerging Tech Trans LLC Setting up multiple channels within ETTDirectControl ...