www.ets-lindgren.com
6505 Shielding Effectiveness Test Kit
41
Frequency
MSA 65-5
Minimum
Attenuation for
Magnetic Field
Approximate
Wattage
Required to
meet MSA 65-6
attenuation
specifications
Maximum
Shielding
(Magnetic Field)
that can be
measured with
this setup
Band 3
400
98
10
110
500
99
15
110
600
99
30
120
700
99
50
120
800
100
40
120
900
100
25
120
1000
100
40
120
Band 4
1 MHz
100
40
120
2
100
25
120
4
100
20
120
6
100
15
120
8
100
15
120
10
100
15
120
15
100
20
120
20
100
25
120
25
100
25
120
30
100
30
120
Содержание 6502
Страница 1: ...6500 Series Loop Antennas User Manual ...
Страница 6: ...vi www ets lindgren com This page intentionally left blank ...
Страница 12: ...12 Introduction www ets lindgren com This page intentionally left blank ...
Страница 18: ...18 Specifications www ets lindgren com This page intentionally left blank ...
Страница 22: ...22 Mounting Instructions www ets lindgren com This page intentionally left blank ...
Страница 26: ...26 Operation www ets lindgren com This page intentionally left blank ...
Страница 27: ...www ets lindgren com Typical Data 27 6 0 Typical Data Model 6502 ...
Страница 28: ...28 Typical Data www ets lindgren com Model 6507 ...
Страница 29: ...www ets lindgren com Typical Data 29 Model 6509 ...
Страница 30: ...30 Typical Data www ets lindgren com Model 6511 ...
Страница 31: ...www ets lindgren com Typical Data 31 Model 6512 ...
Страница 32: ...32 Typical Data www ets lindgren com This page intentionally left blank ...
Страница 36: ...36 Warranty www ets lindgren com This page intentionally left blank ...
Страница 39: ...www ets lindgren com 6505 Shielding Effectiveness Test Kit 39 ...
Страница 42: ...42 6505 Shielding Effectiveness Test Kit www ets lindgren com This page intentionally left blank ...
Страница 43: ...www ets lindgren com EC Declaration of Conformity 43 Appendix C EC Declaration of Conformity Model 6502 ...
Страница 44: ...44 EC Declaration of Conformity www ets lindgren com Model 6507 ...